Enhanced dynamic wedge factors at off‐axis points in asymmetric fields
نویسندگان
چکیده
Several recent reports have described methods for calculating enhanced dynamic wedge factors (EDWFs). Many of these reports use the monitor-unit (MU) fraction method to predict EDWFs as a function of field size. Although simple in approach, MU fraction methods do not produce accurate EDWFs in large or asymmetric fields. A recently described technique, based on the MU fraction method works well for large and asymmetric fields, but only when the calculation point is in the center of the field. Other existing methods based on beam-segment superposition do not have this limitation. These beam summation methods, however, are difficult to implement in routine clinical MU calculation schemes. In this paper, we present a simple calculation method that estimates EDWFs at off-axis calculation points in both symmetric and asymmetric fields. Our method, which also is based on the MU fraction method, similarly uses empirically determined field-size corrections but also applies wedged-field profiles to estimate EDWFs that are independent of calculation-point location and field symmetry. EDWF measurements for a variety of field sizes and calculation-point locations for both 6- and 18-MV x-ray beams were performed to validate our calculations and those of our ADAC Pinnacle3 Treatment Planning System. The disagreement between the calculated and measured EDWFs over the useful clinical range of field sizes and calculation-point locations was less than 2%. The worst disagreement was 3% and occurred at a point 8.5 cm from the center of an asymmetric 25 (wedged direction)x20 cm2 60 degrees-wedged field. Detailed comparisons of measurements with calculations and wedge factors obtained from the ADAC Pinnacle3 Treatment Planning System will be presented. In addition, the strengths and weaknesses of this calculation method will be discussed.
منابع مشابه
Dosimetric evaluation of a treatment planning system using pencil beam convolution algorithm for enhanced dynamic wedges with symmetric and asymmetric fields
Background: The dosimetric performance of Eclipse 6.5 three dimensional treatment planning system (3DTPS) is evaluated by comparing the calculated and measured dose in two dimensions following the guide lines of American Association for Physicists in Medicine Task Group 53. Materials and Methods: The calculations were performed by the 3DTPS for symmetric as well as asymmetric fields for standar...
متن کاملQuantification the dosimetric parameters of asymmetric physical wedged-6MV photon beam
Introduction: Physical wedge as a useful tool has been utilized in radiotherapy to modify photon beam shape and intensity such that it distributes dose uniformly in tumor site and reduces hot points. Since during Linac commissioning dosimetric parameters like output factors and lateral dose profiles are measured only for symmetric open and wedged fields, so calculation the par...
متن کاملExperimental Evaluation of Depth Dose by Exit Surface Diode Dosimeters for Off-Axis Wedged Fields in Radiation Therapy
Introduction Evaluation of the delivered dose of externally wedged photon beams by external diode dosimeters during the treatment process requires the estimation of exit surface dose correction factors in various wedge angles and field sizes. Materials and Methods A system of absorbed dose evaluation, using p-type diode dosimeters placed on the exit surface of a phantom, was characterized for e...
متن کاملCharacterization of Wedge Factors and Dose Distributions in Radiotherapy with Symmetric and Asymmetric Physical Wedged Beams of 6 MV Photon Beam
Introduction: Physical wedge by modify photon beam shape and intensity has been utilized in radiotherapy to obtain uniformly dose distribution in tumor site with reduced hot spots. Calculation of dosimetric parameters for both symmetric and asymmetric wedged fields is proved necessary during linear accelerator (Linac) commissioning. The present study aimed to achieve o...
متن کاملA spreadsheet solution for off‐axis, noncentral enhanced dynamic wedge factors
Enhanced dynamic wedges can be very useful in increasing radiation therapy treatment efficiency. A drawback to their use is finding a factor analogous to a conventional wedge factor to calculate the dose to a point in a dynamically wedged field. A formalism has previously been published [J. Gibbons, Med. Phys. 25, 1411-1418 (1998)] to determine this factor at the central point of a dynamically ...