Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay
نویسندگان
چکیده
Vitamin D status in different populations relies on accurate measurement of total serum 25-hydroxyvitamin D [25(OH)D] concentrations [i.e., 25(OH)D3 and 25(OH)D2]. This study evaluated agreement between the ADVIA Centaur Vitamin D Total assay for 25(OH)D testing (traceable to the NIST-Ghent reference method procedure) and a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for various populations with different levels of vitamin D binding protein (DBP). Total serum 25(OH)D concentrations were measured for 36 pregnant women, 40 hemodialysis patients, and 30 samples (DBP-spiked or not) from healthy subjects. ELISA measured DBP levels. The mean serum DBP concentrations were higher for pregnancy (415 μg/mL) and lower for hemodialysis subjects (198 μg/mL) than for healthy subjects and were highest for spiked serum (545 μg/mL). The average bias between the ADVIA Centaur assay and the LC-MS/MS method was -1.4% (healthy), -6.1% (pregnancy), and 4.4% (hemodialysis). The slightly greater bias for samples from some pregnancy and hemodialysis subjects with serum DBP levels outside of the normal healthy range fell within a clinically acceptable range-reflected by analysis of their low-range (≤136 μg/mL), medium-range (137-559 μg/mL), and high-range (≥560 μg/mL) DBP groups. Thus, the ADVIA Centaur Vitamin D Total assay demonstrates acceptable performance compared with an LC-MS/MS method for populations containing different amounts of DBP.
منابع مشابه
Erratum to “Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay”
[This corrects the article DOI: 10.1155/2014/691679.].
متن کاملPerformance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 Total 25-OH Vitamin D assays.
OBJECTIVES To evaluate the newly developed Roche MODULAR Analytics E170 Total Vitamin D and the Siemens ADVIA Centaur Vitamin D Total assays. MATERIALS AND METHODS Assays were evaluated using the Clinical and Laboratory Standards Institute protocols. Split patient samples were compared with LC-MS/MS and DiaSorin LIAISON assays (n=79 including 15 specimens with detectable endogenous 25-OH vita...
متن کاملPerformance evaluation of two immunoassays for 25-hydroxyvitamin D
Although immunoassays in measuring 25-hydroxyvitamin D [25(OH)D] have been improved recently, relatively large differences are still seen between results of 25(OH)D measured by immunoassays and by liquid chromatography-tandem mass spectrometry (LC-MS/MS). In the present studies, we compared two immunoassays with LC-MS/MS for measuring 25(OH)D concentrations. Concentrations of 25-hydroxyvitamin ...
متن کاملInfluence of the immunoassay used in measurement of serum vitamin D levels.
INTRODUCTION Serum 25-hydroxyvitamin D [25(OH)D] levels are the best indicator of vitamin D levels in the body. Precision, reproducibility, and lack of standardization are the main problems in such measurements. The aim of this study was to compare the 25(OH)D levels measured using Elecsys Vitamin D Total (Roche) and ADVIA Centaur Vitamin D Total (Siemens). MATERIAL AND METHODS 25(OH)D levels...
متن کاملMulticenter comparison study of current methods to measure 25-hydroxyvitamin D in serum.
OBJECTIVES Measurement of serum 25-hydroxyvitamin D [25(OH)D] is generally considered to be a reliable indicator of vitamin D status. The recent increase in diversity of 25(OH)D assays prompted us to evaluate the performance of chromatographic methods (two in-house ID-LC-MS/MS and HPLC (ClinRep, Recipe)), a protein binding method (Cobas-25(OH)D-total, Roche) and immunochemical methods (Liaison ...