OBJECTIVES To determine risk factors for revision in patients implanted with a commonly used metal on metal (MoM) hip replacement. DESIGN Retrospective cohort study in combination with a prospective national retrieval study (Northern Retrieval Registry (NRR)). SETTING Combined orthopaedic unit in combination with the NRR. PARTICIPANTS All patients implanted with a DePuy Pinnacle MoM hip prostheses by the 2 senior authors were invited to attend for a review which included clinical examination, blood metal ion measurements, radiographs and targeted imaging. Explanted components underwent wear analysis using validated methodology and these results were compared with those obtained from the NRR. RESULTS 489 MoM Pinnacle hips were implanted into 434 patients (243 females and 191 males). Of these, 352 patients attended the MoM recall clinics. 64 patients had died during the study period. For the purposes of survival analysis, non-attendees were assumed to have well-functioning prostheses. The mean follow-up of the cohort as a whole was 89 months. 71 hips were revised. Prosthetic survival for the whole cohort was 83.6% (79.9-87.3) at 9 years. The majority of explanted devices exhibited signs of taper junction failure. Risk factors for revision were bilateral MoM prostheses, smaller Pinnacle liners, and implantation in 2006 and later years. A significant number of devices were found to be manufactured out of their specifications. This was confirmed with analysis of the wider data set from the NRR. CONCLUSIONS This device was found to have an unacceptably high revision rate. Bilateral prostheses, those implanted into female patients and devices implanted in later years were found to be at greater risk. A significant number of explanted components were found to be manufactured with bearing diameters outside of the manufacturer's stated tolerances. Our findings highlight the clinical importance of hitherto unrecognised variations in device production.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Total Hip Replacement Revision in a Single Brand Small Cementless Stem – Our Experience after the Findings of the National Joint Registry

Background: Cementless total hip replacement is the common THR performed in England, Wales, Northern Irelandand the Isle of Man. The Corail stem is the most popular cementless implant and has a ODEP 10A rating. Review ofits performance in the registry identified an increase rate of revision amongst the smaller stem sizes. However, claritywas not provided on the explanation for this finding. We ...

متن کامل

Risk factors for failure of the 36 mm metal- on-metal Pinnacle total hip arthroplasty system A RETROSPECTIVE SINGLE-CENTRE COHORT STUDY

Patients and Methods We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multi...

متن کامل

Risk factors for failure of the 36 mm metal-on-metal Pinnacle total hip arthroplasty system

AIMS To determine ten-year failure rates following 36 mm metal-on-metal (MoM) Pinnacle total hip arthroplasty (THA), and identify predictors of failure. PATIENTS AND METHODS We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cau...

متن کامل

Adverse reactions to metal debris occur with all types of hip replacement not just metal-on-metal hips: a retrospective observational study of 3340 revisions for adverse reactions to metal debris from the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man

BACKGROUND Adverse reactions to metal debris (ARMD) have resulted in the high short-term failure rates observed with metal-on-metal hip replacements. ARMD has recently been reported in non-metal-on-metal total hip replacements (non-MoM THRs) in a number of small cohort studies. However the true magnitude of this complication in non-MoM THRs remains unknown. We used a nationwide database to dete...

متن کامل

The clinical implications of elevated blood metal ion concentrations in asymptomatic patients with MoM hip resurfacings: a cohort study

OBJECTIVE To determine whether elevated blood cobalt (Co) concentrations are associated with early failure of metal-on-metal (MoM) hip resurfacings secondary to adverse reaction to metal debris (ARMD). DESIGN Cohort study. SETTING Single centre orthopaedic unit. PARTICIPANTS Following the identification of complications potentially related to metal wear debris, a blood metal ion screening...