Single-Level Rigid Fixation Combined with Coflex: A Biomechanical Study

نویسندگان

  • Wu Che
  • Qian Chen
  • Yi-Qun Ma
  • Yun-Qi Jiang
  • Wei Yuan
  • Xiao-Gang Zhou
  • Xi-Lei Li
  • Jian Dong
چکیده

BACKGROUND The purpose of this biomechanical in vitro study was to compare the kinematics and intradiscal pressure achieved with 2 methods: L4-L5 pedicle screw-rod fixation (PSRF) with an upper L3-L4 Coflex device and L4-L5 PSRF alone. The results were used to characterize the biomechanics of the topping-off operation with a Coflex device for the lumbar motion segment adjacent to single-level rigid fixation. MATERIAL/METHODS Six human cadaveric spine specimens were biomechanically tested in vitro (6 males, 0 females). The 3-dimensional specimen motion in response to applied loads during flexibility tests was determined. Loads were applied along anatomic axes to induce flexion-extension, lateral bending, and axial rotation. All specimens were first studied with intact lumbar motion segments, then with L4-L5 PSRF alone, and finally with L4-L5 PSRF with an upper L3-L4 Coflex device. A non-paired comparison of the 3 configurations under 3 different conditions was made. RESULTS PSRF, with or without a Coflex device, significantly increased the range of motion (ROM) in the upper adjacent motion segments in all directions of loading. The intradiscal pressure (IDP) changed slightly. A correlation analysis showed that the ROM and IDP are significantly positively correlated. The application of the upper motion segment of the Coflex device provided greater stability in all directions of motion than did PSRF alone, particularly for extension (p<0.05), while use of a Coflex device did not significantly decrease the IDP compared with PSRF alone (p>0.05). CONCLUSIONS These results suggest that L4-L5 PSRF with an L3-L4 Coflex device is more stable than L4-L5 PSRF alone. PSRF with an upper Coflex device is a promising alternative to PSRF alone. Based on these biomechanical tests, it might be considered a protective method to prevent adjacent segment degeneration (ASD), although some limitations with this in vitro study must be addressed in the future.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Biomechanical effect after Coflex and Coflex rivet implantation for segmental instability at surgical and adjacent segments: a finite element analysis.

The Coflex device may provide stability to the surgical segment in extension but does not restore stability in other motion. Recently, a modified version called the Coflex rivet has been developed. The effects of Coflex and Coflex rivet implantation on the adjacent segments are still not clear; therefore, the purpose of this study was to investigate the biomechanical differences between Coflex ...

متن کامل

Biomechanical comparison of an interspinous device and a rigid stabilization on lumbar adjacent segment range of motion.

PURPOSE OF THE STUDY Decompression surgery with or without fusion is the gold standard treatment of lumbar spinal stenosis, but adjacent segment degeneration has been reported as a long-term complication after fusion. This led to the development of dynamic implants like the interspinous devices. They are supposed to limit extension and expand the spinal canal at the symptomatic level, but with ...

متن کامل

Biomechanics of a Posterior Lumbar Motion Stabilizing Device: In Vitro Comparison to Intact and Fused Conditions.

STUDY DESIGN Nondestructive flexibility tests were performed in vitro, comparing multiple conditions of fixation in a single group of specimens. OBJECTIVE To compare the biomechanical behavior of the lumbar spine in the intact condition, after implanting a novel motion stabilizer, and after implanting a rigid fixator. SUMMARY OF BACKGROUND DATA Two specific scenarios that may benefit from d...

متن کامل

Finite Element Analysis of the Biomechanical Effect of Coflex™ on the Lumbar Spine

OBJECTIVE The biomechanical properties of the Coflex™ (Paradigm Spine, NY, USA), a device designed to provide dynamic stabilization without lumbar fusion, have not been clearly defined. The purpose of this study was to determine the efficacy and biomechanical effect of Coflex™ using finite element model (FEM). METHODS A 3D geometric model of the L3-L5 was created by integrating computerized t...

متن کامل

Biomechanical Analysis of the Fixation System for T-Shaped Acetabular Fracture

This study aims to evaluate the biomechanical mechanism of fixation systems in the most frequent T-shaped acetabular fracture using finite element method. The treatment of acetabular fractures was based on extensive clinical experience. Three commonly accepted rigid fixation methods (double column reconstruction plates (P × 2), anterior column plate combined with posterior column screws (P + PS...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 22  شماره 

صفحات  -

تاریخ انتشار 2016