Results of the HepZero study comparing heparin-grafted membrane and standard care show that heparin-grafted dialyzer is safe and easy to use for heparin-free dialysis.
نویسندگان
چکیده
Heparin is used to prevent clotting during hemodialysis, but heparin-free hemodialysis is sometimes needed to decrease the risk of bleeding. The HepZero study is a randomized, multicenter international controlled open-label trial comparing no-heparin hemodialysis strategies designed to assess non-inferiority of a heparin grafted dialyzer (NCT01318486). A total of 251 maintenance hemodialysis patients at increased risk of hemorrhage were randomly allocated for up to three heparin-free hemodialysis sessions using a heparin-grafted dialyzer or the center standard-of-care consisting of regular saline flushes or pre-dilution. The first heparin-free hemodialysis session was considered successful when there was neither complete occlusion of air traps or dialyzer, nor additional saline flushes, changes of dialyzer or bloodlines, or premature termination. The current standard-of-care resulted in high failure rates (50%). The success rate in the heparin-grafted membrane arm was significantly higher than in the control group (68.5% versus 50.4%), which was consistent for both standard-of-care modalities. The absolute difference between the heparin-grafted membrane and the controls was 18.2%, with a lower bound of the 90% confidence interval equal to plus 7.9%. The hypothesis of the non-inferiority at the minus 15% level was accepted, although superiority at the plus 15% level was not reached. Thus, use of a heparin-grafted membrane is a safe, helpful, and easy-to-use method for heparin-free hemodialysis in patients at increased risk of hemorrhage.
منابع مشابه
Rationale and design of the HepZero study: a prospective, multicenter, international, open, randomized, controlled clinical study with parallel groups comparing heparin-free dialysis with heparin-coated dialysis membrane (Evodial) versus standard care: study protocol for a randomized controlled trial
BACKGROUND Anticoagulation for chronic dialysis patients with contraindications to heparin administration is challenging. Current guidelines state that in patients with increased bleeding risks, strategies that can induce systemic anticoagulation should be avoided. Heparin-free dialysis using intermittent saline flushes is widely adopted as the method of choice for patients at risk of bleeding,...
متن کاملAvoidance of systemic anticoagulation during intermittent haemodialysis with heparin-grafted polyacrilonitrile membrane and citrate-enriched dialysate: a retrospective cohort study
BACKGROUND Since October 2010, the combination of a heparin-grafted polyacrilonitrile (AN69ST) membrane with a 0.80 mmol/L citric acid-containing dialysate is routinely used in our centre for intermittent haemodialysis, without systemic anticoagulation, in critically ill patients with increased bleeding risk. The primary outcome of this retrospective cohort study was to assess the development o...
متن کاملبررسی مقایسهای تاثیر 2 روش تجویز هپارین به صورت مداوم و متناوب بر روی زمان ترومبوپلاستین نسبی
Comparative Study of Continuous and Intermittent Administration of Heparin on APTT (Activated Partial Thromboplastin Time) I II *A. Adib, MD B. Fatourchi, MD Abstract There are many different doses and protocols of heparin injection for treatment or prevention of many thrombotic diseases. The present interventional experimental(Clinical trial) study was conducted to compare the effect of contin...
متن کاملThe clinical evaluation of low-dose heparin in haemodialysis: a prospective study using the heparin-coated AN69 ST membrane.
BACKGROUND The AN69 ST haemodialysis membrane, a new membrane resulting from coating polyethyleneimine upon the polyacrylonitrile surface, binds heparin. In patients at risk of bleeding, a pilot study has demonstrated the efficient anticoagulant effect of this heparin-coated membrane. Study design. In chronic haemodialyzed patients, we evaluated whether this anticoagulant effect can be validate...
متن کاملPreparation, properties and preclinical pharmacokinetics of low molecular weight heparin-modified isoliquiritigenin-loaded solid lipid nanoparticle
Low molecular weight heparin-modified isoliquiritigenin-loaded solid lipid nanoparticle (LMWH-ISL-SLN) was developed for injective application. The morphological observation, particle diameter and zeta potential of LMWH-ISL-SLN were characterized using transmission electron microscopy (TEM) and a Malvern Zetasizer. Its entrapment efficiency (EE) and drug loading (DL) were determined by ultracen...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Kidney international
دوره 86 6 شماره
صفحات -
تاریخ انتشار 2014