Radiation dose affected by mammographic composition and breast size: first application of a radiation dose management system for full-field digital mammography in Korean women
نویسندگان
چکیده
BACKGROUND Relative to Western women, Korean women show several differences in breast-related characteristics, including higher rates of dense breasts and small breasts. We investigated how mammographic composition and breast size affect the glandular dose during full-field digital mammography (FFDM) in Korean women using a radiation dose management system. METHODS From June 1 to June 30, 2015, 2120 FFDM images from 560 patients were acquired and mammographic breast composition and breast size were assessed. We analyzed the correlations of patient age, peak kilovoltage (kVp), current (mAs), compressed breast thickness, compression force, mammographic breast composition, and mammographic breast size with the mean glandular dose (MGD) of the breast using a radiation dose management system. The causes of increased radiation were investigated, among patients with radiation doses above the diagnostic reference level (4th quartile, ≥75%). RESULTS The MGD per view of 2120 images was 1.81 ± 0.70 mGy. In multivariate linear regression analysis, age was negatively associated with MGD (p < 0.05). The mAs, kVp, compressed breast thickness, and mammographic breast size were positively associated with MGD (p < 0.05). The "dense" group had a significantly higher MGD than the "non-dense" group (p < 0.05). Patients with radiation dose values above the diagnostic reference value had large breasts of dense composition. CONCLUSIONS Among Korean women, patients with large and dense breasts should be more carefully managed to ensure that a constant radiation dose is maintained.
منابع مشابه
Estimation of Mean Estimation of Mean Glandular Dose in Mammography Examination in some of Radiology Centers in Tehran city
Introduction: the objectives of this research were determination of Mean Glandular Dose in film- screen and digital mammography examinations, estimation of accumulated effective dose in some of mammography centers in Tehran city and estimating cancer risk due to these examinations. Materials and Methods: the clinical data were collected from the 1318 <...
متن کاملارزیابی دز و کیفیت تصویر در آزمون های ماموگرافی در مراکز ماموگرافی شهر تهران
Abstract Background: Breast cancer is one of the most common form of cancer among women. Diagnosis in the early stage of emergence is the most effective way to reduce the mortality caused by this disease. Mammography, the safest and most accurate imaging test for the identification of breast cancer, required high quality images along applying the lowest possible dose to sensitive tissues of the...
متن کاملCalculation of Absorbed Glandular Dose using a FORTRAN Program Based on Monte Carlo X-ray Spectra in Mammography
Introduction: Average glandular dose calculation in mammography with Mo-Rh target-filter and dose calculation for different situations is accurate and fast. Material and Methods: In this research, first of all, x-ray spectra of a Mo target bombarded by a 28 keV electron beam with and without a Rh filter were calculated using the MCNP code. Then, we used the Sobol-Wu parameters to write a FORTRA...
متن کاملDiagnosing breast cancer: an opportunity for innovative engineering.
BACKGROUND In an ancient papyrus, dating back to 2500 BC, an Egyptian named Imhotep, thought to be the first physician and engineer in history, offers a glimpse into the earliest recorded case of breast cancer. Treatment has evolved over the years, from radical mastectomy to radiation by X-rays, and from adjuvant chemotherapy to hormonal therapy. A key challenge in fighting the 'war' on breast ...
متن کاملCalculating Contra-Lateral Breast’s Absorption Dose in Different Radiotherapy Methods of Large Breasts Cancer
Introduction: Breast cancer is one of the most common known cancers in women all around the world. It is the most common cause of death in women after lung cancer. Radiotherapy is a method for therapy of cancer in which certain dose of high energy radiations is used to destruct cancer cells. Radiotherapy is designed so that radiation of eartain does to the target area in body i...
متن کامل