The effects of differential injection sites of cold saline on transpulmonary thermodilution parameter values

نویسندگان

  • Wanjie Yang
  • Qingguo Feng
  • Youzhong An
  • Xuefeng Zhao
  • Kai Wei
  • Chang Li
  • Wei Wang
  • Hongyun Teng
چکیده

AIM To investigate the effects of differential sites for cold saline injection on transpulmonary thermodilution parameter values. METHODS This was a prospective study. Twelve patients were recruited for the following examinations: control condition (injection site at proximal injection end of the Swan-Ganz catheter), proximal end condition (injection site at sheath of the Swan-Ganz catheter), and distal end condition (injection site at PA end of the Swan-Ganz catheter). Sixty measurements were performed for each condition. The cardiac index, global end diastolic volume index (GEDI), and extravascular lung water index for the three different injection sites were recorded from each patient. In addition, the mean transmission time (MTt), downslope time, and area under the curve obtained from PiCCO-VoLEF-Win software were compared among different groups. RESULTS There were no differences in cardiac index and extravascular lung water index values among the three conditions (P>0.05). There were no differences in GEDI between the proximal end condition and control condition (P>0.05), while the GEDI was significantly lower for the distal end condition (493.33±254.65 mL/m(2)) than for the control condition (645.53±234.46 mL/m(2)) (P<0.05) and proximal end condition (717.96±321.63 mL/m(2)) (P<0.01). There were no differences in downslope time and area under the curve among the three conditions (P>0.05). There were no differences in MTt between the proximal end condition and control condition (P>0.05), while the MTt was significantly lower for distal end condition (40.22±16.37 seconds) than for the control condition (42.91±17.93 seconds) (P<0.05) and proximal end condition (47.16±16.64 seconds) (P<0.01). CONCLUSION The differential sites for cold saline injection impacted transpulmonary thermodilution parameter values.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Is the recommendation not to use rifampin plus pyrazinamide for latent tuberculosis treatment always imperative?

saline solution bolus via a central venous catheter, a thermistor-tipped arterial catheter (usually femoral) is connected to a specific monitor (PiCCOplus; Pulsion Medical Systems; Munich, Germany), which records the downstream temperature changes, ie, a transpulmonary thermodilution curve. The mathematical analysis of the curve allows the computation of cardiac output, heart blood volume (glob...

متن کامل

Precision of the transpulmonary thermodilution measurements

INTRODUCTION We wanted to determine the number of cold bolus injections that are necessary for achieving an acceptable level of precision for measuring cardiac index (CI), indexed global end-diastolic volume (GEDVi) and indexed extravascular lung water (EVLWi) by transpulmonary thermodilution. METHODS We included 91 hemodynamically stable patients (age 59 (25% to 75% interquartile range: 39 t...

متن کامل

Transpulmonary thermodilution: advantages and limits

BACKGROUND For complex patients in the intensive care unit or in the operating room, many questions regarding their haemodynamic management cannot be answered with simple clinical examination. In particular, arterial pressure allows only a rough estimation of cardiac output. Transpulmonary thermodilution is a technique that provides a full haemodynamic assessment through cardiac output and othe...

متن کامل

Transpulmonary thermodilution assessments: precise measurements require a precise procedure

When incorporating the values of a hemodynamic parameter into the care of patients, the precision of the measurement method should always be considered. A prospective analysis in the previous issue of Critical Care showed that the precision of transpulmonary thermodilution (TPTD) allows for reliable mean values if a standardised procedure is used. The present finding has a physiological basis, ...

متن کامل

Validation of extravascular lung water measurement by single transpulmonary thermodilution: human autopsy study

INTRODUCTION Gravimetric validation of single-indicator extravascular lung water (EVLW) and normal EVLW values has not been well studied in humans thus far. The aims of this study were (1) to validate the accuracy of EVLW measurement by single transpulmonary thermodilution with postmortem lung weight measurement in humans and (2) to define the statistically normal EVLW values. METHODS We eval...

متن کامل