Towards real-time cardiovascular magnetic resonance-guided transarterial aortic valve implantation: In vitro evaluation and modification of existing devices
نویسندگان
چکیده
BACKGROUND Cardiovascular magnetic resonance (CMR) is considered an attractive alternative for guiding transarterial aortic valve implantation (TAVI) featuring unlimited scan plane orientation and unsurpassed soft-tissue contrast with simultaneous device visualization. We sought to evaluate the CMR characteristics of both currently commercially available transcatheter heart valves (Edwards SAPIEN™, Medtronic CoreValve®) including their dedicated delivery devices and of a custom-built, CMR-compatible delivery device for the Medtronic CoreValve® prosthesis as an initial step towards real-time CMR-guided TAVI. METHODS The devices were systematically examined in phantom models on a 1.5-Tesla scanner using high-resolution T1-weighted 3D FLASH, real-time TrueFISP and flow-sensitive phase-contrast sequences. Images were analyzed for device visualization quality, device-related susceptibility artifacts, and radiofrequency signal shielding. RESULTS CMR revealed major susceptibility artifacts for the two commercial delivery devices caused by considerable metal braiding and precluding in vivo application. The stainless steel-based Edwards SAPIEN™ prosthesis was also regarded not suitable for CMR-guided TAVI due to susceptibility artifacts exceeding the valve's dimensions and hindering an exact placement. In contrast, the nitinol-based Medtronic CoreValve® prosthesis was excellently visualized with delineation even of small details and, thus, regarded suitable for CMR-guided TAVI, particularly since reengineering of its delivery device toward CMR-compatibility resulted in artifact elimination and excellent visualization during catheter movement and valve deployment on real-time TrueFISP imaging. Reliable flow measurements could be performed for both stent-valves after deployment using phase-contrast sequences. CONCLUSIONS The present study shows that the Medtronic CoreValve® prosthesis is potentially suited for real-time CMR-guided placement in vivo after suggested design modifications of the delivery system.
منابع مشابه
Real-time MR-guided transarterial aortic valve implantation (tavi): in vivo evaluation in swine
Objective Transcatheter, transarterial aortic valve implantation (TAVI) is rapidly emerging as a promising new treatment option for patients with severe symptomatic aortic valve stenosis who are considered at high or prohibitive surgical risk. Envisioning real-time MR guidance of the TAVI procedure, the objective of this study was the systematical in vitro evaluation of the MR imaging character...
متن کاملTowards real-time cardiovascular magnetic resonance guided transarterial CoreValve implantation: in vivo evaluation in swine
BACKGROUND Real-time cardiovascular magnetic resonance (rtCMR) is considered attractive for guiding TAVI. Owing to an unlimited scan plane orientation and an unsurpassed soft-tissue contrast with simultaneous device visualization, rtCMR is presumed to allow safe device navigation and to offer optimal orientation for precise axial positioning. We sought to evaluate the preclinical feasibility of...
متن کاملReal-time magnetic resonance imaging-guided transarterial aortic valve implantation: in vivo evaluation in swine.
متن کامل
Magnetic resonance-guided cardiac interventions using magnetic resonance-compatible devices: a preclinical study and first-in-man congenital interventions.
BACKGROUND Percutaneous cardiac interventions are currently performed under x-ray guidance. Magnetic resonance imaging (MRI) has been used to guide intravascular interventions in the past, but mainly in animals. Translation of MR-guided interventions into humans has been limited by the lack of MR-compatible and safe equipment, such as MR guide wires with mechanical characteristics similar to st...
متن کاملQuantitative analysis of post-TAVI aortic regurgitation with cardiovascular magnetic resonance and the relationship to transthoracic echocardiography
Background Transcatheter Aortic Valve Implantation (TAVI) is increasingly used to treat patients with severe aortic stenosis at high surgical risk. The severity of post-implantation valvular or paravalvular regurgitation has been shown to adversely affect patient outcome. The aim of the study was to assess the prevalence and severity of aortic regurgitation (AR) at 6 months post-TAVI using card...
متن کامل