Intrinsic structural failure of polyester (Dacron) vascular grafts. A general review.
نویسندگان
چکیده
Intrinsic structural failure of Dacron prostheses is a late exceptional complication, resulting from a loss of structural integrity of the graft. The authors report six cases of non-anastomotic false aneurysms in the mid-portion of a vascular Dacron graft, observed at a mean of 12 years after insertion. It concerns four femoro-popliteal bypass grafts, one cross-over graft and a branch of a bifurcated aorto-bifemoral graft, implanted between 1980 and 1990. This represents 0.2% of all vascular Dacron grafts implanted in authors' department since 1980. The degenerated prosthesis was excised, and a new bypass graft was inserted. In three cases, histological analysis revealed a foreign body giant cell reaction against fragmented Dacron fibers. In none of the cases there was evidence of graft infection. The authors discuss the evidence and etiopathogeny of this late, unusual complication, inherent to the Dacron graft material. The most probable causative factor is material fatigue, leading to gradual breakdown and fragmentation of individual fibers, and subsequent biodegradation of the basic material. Such an intrinsic weakness of prosthetic fabric has only be observed in first and second generation Dacron grafts.
منابع مشابه
Dacron® vs. PTFE as bypass materials in peripheral vascular surgery – systematic review and meta-analysis
BACKGROUND In peripheral vascular bypass surgery different synthetic materials are available for bypass grafting. It is unclear which of the two commonly used materials, polytetrafluoroethylene (PTFE) or polyester (Dacron(R)) grafts, is to be preferred. Thus, the aim of this meta-analysis and systematic review was to compare the effectiveness of these two prosthetic bypass materials (Dacron and...
متن کاملA contemporary meta-analysis of Dacron versus polytetrafluoroethylene grafts for femoropopliteal bypass grafting.
BACKGROUND The present study provides a contemporary and comprehensive summation of midterm patency rates of polyester (Dacron) or polytetrafluoroethylene (PTFE) grafts in femoropopliteal bypass grafting based on a meta-analysis consisting only of randomized controlled trials. METHODS MEDLINE, EMBASE, and the Cochrane Central Register of Controlled Trials were searched to identify all randomi...
متن کاملNon-Anastomotic Rupture of a Woven Dacron Graft in the Descending Thoracic Aorta Treated with Endovascular Stent Grafting
The intrinsic structural failure of a Dacron graft resulting from the loss of structural integrity of the graft fabric can cause late graft complications. Late non-anastomotic rupture has traditionally been treated surgically via open thoracotomy. We report a case of the successful use of thoracic endovascular repair to treat a Dacron graft rupture in the descending aorta. The rupture occurred ...
متن کاملVASCULAR ACCESS FOR HEMODIALYSIS: OUR EXPERIENCE WITH 3337 CASES
Over a seven year period from 1982 to 1989,3337 conduits were created in 3137 patients with end stage renal disease (ESRD) as access for chronic dialysis. These included 2690 side-to-side arteriovenous fistulae (A VF), 168 end-to-side A VFs, 10 autogenous vein grafts, 51 homogenous frozen vein grafts, 109 polytetrafluoroethylene (PTFE) grafts, 209 emergency external arteriovenous shunts, t...
متن کاملDacron graft aneurysm with dissection
Dacron grafts have been used as a conduit for large caliber arteries for many years successfully. However, these grafts can undergo complications such as aneurysm formation, rupture, and failure. Evaluation of these complications are of paramount importance because of its tendency to rupture and cause death. Imaging plays an important role in identifying and monitoring of these complications, a...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Acta chirurgica Belgica
دوره 105 3 شماره
صفحات -