Total Power Radiometer for Medical Sensor Applications Using Matched and Mismatched Noise Sources
نویسندگان
چکیده
This paper presents a simple total power radiometer to noninvasively measure the temperature of the human body. The proposed 3-GHz radiometer consists of an antenna collecting the noise power generated by a target, a low-noise and high-gain receiver amplifying the noise power, and a detector converting the noise power to voltage. A single-pole-triple-throw (SP3T) switch is placed between the antenna and the receiver, while a personal computer is used to control the SP3T switch, collect and process the data such as detector output voltages and physical temperatures of the reference noise sources and the target. The fabricated radiometer shows a good performance agreement with a thermometer in the temperature measurement of water from 25.0 to 43.1 °C. For the accurate prediction of the target temperature, the radiometer is calibrated adaptively to the environment and radiometer variations. For this purpose, two reference noise sources (hot and cold) are proposed using matched and mismatched resistors at room temperature. These resistor-based noise sources offer a reliable performance without complex temperature control systems. Furthermore, they can be easily calibrated in real time by periodically measuring the physical temperatures of the resistors. In addition, the logarithmic detector with wide dynamic range is adopted and logarithmically-fitted based on the measurement results instead of linear approximation, which reduces the error caused by the limited dynamic range of resistor-based noise sources. In order to further increase the accuracy, the performance imbalances between ports in the SP3T switch are also taken into account by employing offsets in the radiometer output voltages.
منابع مشابه
A W-band Simultaneously Matched Power and Noise Low Noise Amplifier Using CMOS 0.13µm
A complete procedure for the design of W-band low noise amplifier in MMIC technology is presented. The design is based on a simultaneously power and noise matched technique. For implementing the method, scalable bilateral transistor model parameters should be first extracted. The model is also used for transmission line utilized in the amplifier circuit. In the presented method, input/output ma...
متن کاملEvaluation of Uncertainties of the Null-Balanced Total-Power Radiometer System NCSl
National Institute of Standards and Technology, Boulder. CO 80303-3328 Standard unrertaintjes are evaluated for the nuU-balarced, total-power, heicrodyned radiometer system with a switched input that was recently developed U NIST lo calibrate thermal noise sources. Eighi significant sources of uncertainty due tw sysletoatic effects arc iiicnlirtcd. two attributable to the two noise standards, a...
متن کاملDesign of High Gain, High Reverse Isolation and High Input Matched Narrowband LNA for GPS L1 Band Applications Using 0.18µm Technology
Design of Global Positioning System (GPS) receiver with a low noise amplifier (LNA) in the front end remains a major design requirement for the success of modern day navigation and communication system. Any LNA is expected to meet the requirements like its ability to add the least amount of noise while providing sufficient gain, perfect input and output matching, and high linearity. However, mo...
متن کاملDesign of Medical Radiometer Front-end for Improved Performance.
We have investigated the possibility of building a singleband Dicke radiometer that is inexpensive, small-sized, stable, highly sensitive, and which consists of readily available microwave components. The selected frequency band is at 3.25-3.75 GHz which provides a reasonable compromise between spatial resolution (antenna size) and sensing depth for radiometry applications in lossy tissue. Fore...
متن کاملInitial Plan for a Precision Temperature Sensor
Precision microwave radiometry requires accurate internal calibration sources. A microwave terminator is one such internal source, and, under the assumption of a high quality match, produces a noise power directly proportional to the physical temperature of the terminator. Precise measurement of this temperature is required in order to provide a quality calibration standard. It is desirable for...
متن کامل