CTXA Hip—An Extension of Classical DXA Measurements Using Quantitative CT
نویسندگان
چکیده
Bone mineral density (BMD) estimates for the proximal femur using Dual Energy X-ray Absorptiometry (DXA) are currently considered the standard for making a diagnosis of osteoporosis in an individual patient using BMD alone. We have compared BMD results from a commercial Quantitative CT (QCT) BMD analysis system, "CTXA Hip", which provides clinical data for the proximal femur, to results from DXA. We have also used CTXA Hip to determine cortical and trabecular contributions to total BMD. Sixty-nine patients were scanned using 3D QCT and DXA. CTXA Hip BMD measurements for Total Hip and Femoral Neck were compared to DXA results. Twenty-two women were scanned at 0, 1, 2 years and CTXA Hip and DXA results analyzed for long-term reproducibility. Long-term reproducibility calculated as root-mean-square averages of SDs in vivo was 0.012 g/cm2 (CV = 1.8%) for CTXA Total Hip and 0.011 g/cm2 (CV = 2.0%) for CTXA Femoral Neck compared to 0.014 g/cm2 (CV = 2.0%) and 0.016 g/cm2 (CV = 2.7%), respectively, for DXA. The correlation of Total Hip BMD CTXA vs. DXA was R = 0.97 and for Femoral Neck was R = 0.95 (SEE 0.044 g/cm2 in both cases). Cortical bone comprised 62±5% (mean ± SD) of total hipbone mass in osteoporotic women. CTXA Hip provides substantially the same clinical information as conventional DXA and in addition provides estimates of BMD in separate cortical and trabecular bone compartments, which may be useful in evaluation of bone strength.
منابع مشابه
A Physician’s Guide to QCT BMD Testing
Today, QCT is one of three methods, including DXA, cited by the National Osteoporosis Foundation as safe and effective for the evaluation of BMD[3]. The QCT system comprises a CT scanner calibration device and a software application on a standalone Windows PC networked to the CT scanner. At the hip, QCT produces BMD T-Score measurements that are clinically equivalent to DXA measurements[5], but...
متن کاملCorrelation between Periacetabular Bone Density Changes and Stress Behaviour after Total Hip Arthroplasty
Introduction Little is know about periacetabular bone remodeling after insertion of the acetabular cup in total hip Arthroplasty. Dual energy X-ray absorptiometry (DXA) is commonly used for quantitative assessment of bone remodeling after total hip arthroplasty (THA). DXA is limited by low resolution and a lack of three-dimensional information. Computed tomography (CT) is a standardized radiolo...
متن کاملHip structural analysis: a comparison of DXA with CT in postmenopausal Japanese women
Geometry of the proximal femur is one determinant of fracture risk, and can be analyzed by a simple method using dual-energy X-ray absorptiometry (DXA). The aim of the present study was to investigate the accuracy of hip structural analysis (HSA) using clinical data in postmenopausal Japanese women. A total of 184 postmenopausal women aged 51-88 years (mean, 70.5 ± 8.7 years) who underwent arti...
متن کاملPrediction of total and visceral fat contents using anthropometric measures of adiposity in women.
BACKGROUND Although techniques such as dual energy xray absorptiometry (DXA) and quantitative CT are available to estimate global and regional adiposity, anthropometric measurements are often used to detect adiposity in clinical practice. OBJECTIVES To assess the association between the anthropometric measurements of obesity with total and regional fat mass determined by DXA. DESIGN A cross...
متن کاملClinical use of quantitative computed tomography and peripheral quantitative computed tomography in the management of osteoporosis in adults: the 2007 ISCD Official Positions.
The International Society for Clinical Densitometry (ISCD) has developed Official Positions for the clinical use of dual-energy X-ray absorptiometry (DXA) and non-DXA technologies. While only DXA can be used for diagnostic classification according to criteria established by the World Health Organization, DXA and some other technologies may predict fracture risk and be used to monitor skeletal c...