Development of a new in-water-phantom graphite calorimeter for the measurement of absorbed dose to water in medium energy x-ray beams
نویسندگان
چکیده
This paper summarizes the progress on the construction of a graphite calorimeter for the measurement of absorbed dose to water Dw in medium-energy filtered x-ray qualities, as part of the EMRPfunded project “Metrology for radiotherapy using complex radiation fields” (MetrExtRT, http://radiotherapyemrp.eu/). The design and the construction of a new primary standard for absorbed dose to water (Dw) is shown, together with a presentation of the preliminary estimates of conversion factors, calculated by Monte Carlo using the EGSnrc code, and heat-transport calculations resulting from COMSOL simulations. The vacuum-gap-insulated calorimeter core has a thickness of 2 mm and a diameter of 20 mm. The graphite calorimeter is embedded in a PMMA waterproof sleeve and can be positioned at the desired measurement depth in a water phantom. Once completed, the new calorimeter will participate in an international comparison with other existing standards of absorbed dose to water for medium energy x-rays.
منابع مشابه
The Graphite Calorimeter as a Standard of Absorbed Dose for Cobalt-60 Gamma Radiation
Absorbed dose to water in a cobalt-60 gamma-ray beam has been determined using a thick-walled graphite ionization chamber. The chamber was calibrated in a graphite phantom against a graphite calorimeter, and the graphite calibration factor was converted to a water calibration factor using published energy absorption coefficient ratios and a measured replacement factor. Comparisons between the g...
متن کاملMeasuring the temperature dependence of refractive index in water phantom for precise monitoring of the absorbed dose by radiation therapy techniques
In this paper, an experimental setup is used to measure the temperature dependence of the refractive index of a water phantom. In radiation calorimetry by laser beams and interferometric systems, the amount of change induced by radiation within the phantom can be accurately measured. Absorption of dose and the resulted change in temperature changes the refractive index within the material. In o...
متن کاملThe UK primary standard calorimeter for photon-beam absorbed dose measurement.
A description is given of the UK primary standard graphite calorimeter system. The calorimeter measures absorbed dose to graphite for photon radiations from 60Co to 19 MV x-rays, and is the basis of the NPL therapy-level absorbed dose to water calibration service. Absorbed dose to graphite from the photon calorimeter has been compared with three other standards: an ionization chamber and cavity...
متن کاملVolumetric modulated arc therapy ( VMAT )
New calorimeters Simon Duane, NPL Calorimetry remains the most fundamental method for the measurement of absorbed dose and is the basis of almost all absorbed dose primary standards worldwide. Absorbed dose to water is realised either directly, by using water as the absorbing medium, or indirectly, by using another medium (usually graphite) and determining a factor to convert dose from that med...
متن کاملComparison of the photon charge between water and solid phantom depending on depth
Background: This study assessed the clinical usefulness of the solid phantom, which may compensate for the disadvantages of the water phantom, by comparing the radiological doses between the two depending on their depths. Materials and Methods: The experimental equipment used was a linear accelerator for medical use, water phantom, solid phantom, Farmer type ion chamber and electrometer. The di...
متن کامل