Multicenter, Randomized, Controlled Trial of 150-J Biphasic Shocks Compared With 200- to 360-J Monophasic Shocks in the Resuscitation of Out-of-Hospital Cardiac Arrest Victims
نویسندگان
چکیده
Background—In the present study, we compared an automatic external defibrillator (AED) that delivers 150-J biphasic shocks with traditional high-energy (200to 360-J) monophasic AEDs. Methods and Results—AEDs were prospectively randomized according to defibrillation waveform on a daily basis in 4 emergency medical services systems. Defibrillation efficacy, survival to hospital admission and discharge, return of spontaneous circulation, and neurological status at discharge (cerebral performance category) were compared. Of 338 patients with out-of-hospital cardiac arrest, 115 had a cardiac etiology, presented with ventricular fibrillation, and were shocked with an AED. The time from the emergency call to the first shock was 8.963.0 (mean6SD) minutes. Conclusions—The 150-J biphasic waveform defibrillated at higher rates, resulting in more patients who achieved a return of spontaneous circulation. Although survival rates to hospital admission and discharge did not differ, discharged patients who had been resuscitated with biphasic shocks were more likely to have good cerebral performance. (Circulation. 2000;102:1780-1787.)
منابع مشابه
BIPHASIC Trial: a randomized comparison of fixed lower versus escalating higher energy levels for defibrillation in out-of-hospital cardiac arrest.
BACKGROUND There is little clear evidence as to the optimal energy levels for initial and subsequent shocks in biphasic waveform defibrillation. The present study compared fixed lower- and escalating higher-energy regimens for out-of-hospital cardiac arrest. METHODS AND RESULTS The Randomized Controlled Trial to Compare Fixed Versus Escalating Energy Regimens for Biphasic Waveform Defibrillat...
متن کاملA Randomized Comparison of Fixed Lower Versus Escalating Higher Energy Levels for Defibrillation in Out-of-Hospital Cardiac Arrest
Background—There is little clear evidence as to the optimal energy levels for initial and subsequent shocks in biphasic waveform defibrillation. The present study compared fixed lowerand escalating higher-energy regimens for out-of-hospital cardiac arrest. Methods and Results—The Randomized Controlled Trial to Compare Fixed Versus Escalating Energy Regimens for Biphasic Waveform Defibrillation ...
متن کاملBiphasic versus monophasic shock waveform for conversion of atrial fibrillation.
Cardioversion of atrial fibrillation (AF) using traditional monophasic shock waveform is unsuccessful in up to 20% of cases, and often requires several shocks of up to 360 J. Based on the success with biphasic shock waveform in converting ventricular fibrillation, it was postulated that biphasic shocks would allow cardioversion with lower energy. In a international multicenter, double-blind, ra...
متن کاملA comparison of biphasic and monophasic waveform defibrillation after prolonged ventricular fibrillation.
STUDY OBJECTIVE To compare the effects of biphasic defibrillation waveforms and conventional monophasic defibrillation waveforms on the success of initial defibrillation, postresuscitation myocardial function, and duration of survival after prolonged duration of untreated ventricular fibrillation (VF), including the effects of epinephrine. DESIGN Prospective, randomized, animal study. SETTI...
متن کاملA multicenter prospective randomized study comparing the efficacy of escalating higher biphasic versus low biphasic energy defibrillations in patients presenting with cardiac arrest in the in-hospital environment
BACKGROUND Biphasic defibrillation has been practiced worldwide for >15 years. Yet, consensus does not exist on the best energy levels for optimal outcomes when used in patients with ventricular fibrillation (VF)/pulseless ventricular tachycardia (VT). METHODS This prospective, randomized, controlled trial of 235 adult cardiac arrest patients with VF/VT was conducted in the emergency and card...
متن کامل