Aneurysm clips: evaluation of magnetic field interactions with an 8.0 T MR system.

نویسندگان

  • A Kangarlu
  • F G Shellock
چکیده

This study was conducted to evaluate magnetic field interactions for aneurysm clips exposed to an 8.0 T magnetic resonance (MR) system. Twenty-six different aneurysm clips were tested for magnetic field translational attraction (deflection angle test) and torque (qualitative assessment method) using previously described techniques. Six of the specific aneurysm clips (i.e. type, model, blade length) made from stainless steel alloy (Perneczky) and Phynox (Yasargil, models FE 748 and FE 750) displayed deflection angles above 45 degrees and torque measurements of +4, indicating that these aneurysm clips maybe unsafe for patients or individuals in an 8.0 T MR environment. The specific aneurysm clips (i.e. type, model, blade length) made from commercially pure titanium (Spetzler), Elgiloy (Sugita), titanium alloy (Yasargil, model FE 750T), and MP35N (Sundt) displayed deflection angles less than 45 degrees and torque that ranged from + 1 to +4. Accordingly, these aneurysm clips are likely to be safe for patients or individuals exposed to an 8.0 T MR system. Depending on the actual dimensions and mass, an aneurysm clip made from Elgiloy may or may not be acceptable for a patient or individual in the 8.0 T MR environment.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Aneurysm clips: evaluation of magnetic field interactions and translational attraction by use of "long-bore" and "short-bore" 3.0-T MR imaging systems.

BACKGROUND AND PURPOSE The use of 3.0-T MR systems is increasing worldwide. We evaluated magnetic field interactions and translational attraction for 32 aneurysm clips in association with exposure to "long-bore" and "short-bore" 3.0-T MR imaging systems. METHODS Thirty-two different aneurysm clips were evaluated in this investigation. Each aneurysm clip was qualitatively evaluated for magneti...

متن کامل

In vitro evaluation of MR imaging issues at 3T for aneurysm clips made from MP35N: Findings and information applied to 155 additional aneurysm clips.

BACKGROUND AND PURPOSE Aneurysm clips need to be tested at 3T to characterize MR imaging concerns, including magnetic field interactions, MR imaging-related heating, and artifacts. Therefore, we evaluated these risks for aneurysm clips. MATERIALS AND METHODS Three different MP35N aneurysm clips (Codman Slim-Line Aneurysm Clip, straight, blade length 25-mm; Codman Slim-Line Aneurysm Clip Graft...

متن کامل

MR imaging of patients with intracranial aneurysm clips.

Four patients with intracranial aneurysm clips made from a variety of alloys were studied without incidence by MR imaging at field strengths ranging from 0.35 to 0.6 T. Knowledge of the type of alloy used in the manufacturing of an aneurysm clip is important in determining whether the clip will or will not deflect in a magnetic field. Ferromagnetic clips show deflection and torque in a magnetic...

متن کامل

Aneurysm clips made of titanium: magnetic characteristics and artifacts in MR.

PURPOSE To evaluate the magnetic characteristics, artifact formation, and implant safety of titanium aneurysm clips for use in MR imaging. METHODS Aneurysm clips made of titanium alloy TiAl6V4 were tested in a magnetometer to determine their magnetic susceptibility and in a 1.5-T MR imager using both a geometric phantom and an animal model. A commercially available alpha-Phynox clip served as...

متن کامل

Assessment of MRI issues at 7 T for 28 implants and other objects.

OBJECTIVE Metallic implants are currently a contraindication for volunteer subjects and patients referred for 7-T examinations because of concerns related to magnetic field interactions and MRI-related heating. Artifacts may also be problematic. Therefore, the purpose of this investigation was to evaluate these MRI issues for 28 implants and other objects in association with a 7-T MR system. ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of magnetic resonance imaging : JMRI

دوره 12 1  شماره 

صفحات  -

تاریخ انتشار 2000