Estimation of uncertainty in three dimensional coordinate measurement by comparison with calibrated points
نویسنده
چکیده
This paper details a method of estimating the uncertainty of dimensional measurement for a three dimensional coordinate measurement machine. An experimental procedure was developed to compare three dimensional coordinate measurements with calibrated reference points. The reference standard used to calibrate these reference points was a fringe counting interferometer with a multilateration like technique employed to establish three dimensional coordinates. This is an extension of the established technique of comparing measured lengths with calibrated lengths. Specifically a distributed coordinate measurement device was tested which consisted of a network of Rotary-Laser Automatic Theodolites (R-LATs), this system is known commercially as indoor GPS (iGPS). The method was found to be practical and was used to estimate that the uncertainty of measurement for the basic iGPS system is approximately +/1 mm at a 95% confidence level throughout a measurement volume of approximately 10 m x 10 m x 1.5 m.
منابع مشابه
Verification of the indoor GPS system, by comparison with calibrated coordinates and by angular reference
This paper details work carried out to verify the dimensional measurement performance of the Indoor GPS (iGPS) system; a network of Rotary-Laser Automatic Theodolites (R-LATs). Initially tests were carried out to determine the angular uncertainties on an individual R-LAT transmitter-receiver pair. A method is presented of determining the uncertainty of dimensional measurement for a three dimens...
متن کاملVerification of the Indoor GPS System by Comparison with Points Calibrated Using a Network of Laser Tracker Measurements
This paper details a method of determining the uncertainty of dimensional measurement for a three dimensional coordinate measurement machine. An experimental procedure was developed to compare three dimensional coordinate measurements with calibrated reference points. The reference standard used to calibrate these reference points was a fringe counting interferometer with the multilateration te...
متن کاملImprovement of Accuracy in Environmental Dosimetry by TLD Cards Using Three-dimensional Calibration Method
Background: The angular dependency of response for TLD cards may cause deviation from its true value on the results of environmental dosimetry, since TLDs may be exposed to radiation at different angles of incidence from the surrounding area.Objective: A 3D setting of TLD cards has been calibrated isotropically in a standard radiation field to evaluate the improvement of the accuracy of meas...
متن کاملمقایسه تفاوت لندمارک های آناتومیک دو تکنیک رادیوگرافی لترال سفالومتری دیجیتال و معمولی به روش ترسیم دستی در جمجمه های انسانی
Background and purpose: Digital radiography has led to many improvements in radiology. Despite many advantages there are different ideas in determining anatomical landmarks which results in some errors in cephalometric analyses. The aim of this study was to determine the degree of identification differences of anatomical landmarks by conventional and digital lateral cephalometric techniques usi...
متن کاملConcept for an Algorithm Testing and Evaluation Program at NIST
F i g u re 1 Coord ina t e measu r ing system—data analysis software analyzes coordinate data, producing surface fits. "any piece of equipment which collects three dimensional coordinates (points), calculates and displays additional information using the measured points." [1]. Figure 1 displays a high level view of how a CMS works. A CMS—such as a coordinate measurement machine, vision system, ...
متن کامل