Tissue equivalence correction for silicon microdosimetry detectors in boron neutron capture therapy.

نویسندگان

  • P D Bradley
  • A B Rosenfeld
چکیده

Reverse-biased silicon p-n junction arrays have been proposed as microdosimetry detectors. The tissue equivalence of such detectors in boron neutron capture therapy (BNCT) is discussed. A comparison of the range-energy relationships of H, He, C, and Li ions in tissue (ICRU-muscle) and silicon is given. A simple geometrical scaling (approximately 0.63) of linear dimensions is required to convert microdosimetric energy deposition measurements performed in silicon to equivalent deposition in tissue. The Monte Carlo technique is used to examine energy deposition for two simple geometrical cases applicable to BNCT.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Performance of silicon microdosimetry detectors in boron neutron capture therapy.

Reverse-biased silicon p-n junction arrays using Silicon-On-Insulator technology have been proposed as microdosimeters. The performance of such detectors in boron neutron capture therapy (BNCT) is discussed. This work provides the first reported measurements using boron-coated silicon diode arrays as microdosimeters in BNCT. Results are in good agreement with measurements with gas proportional ...

متن کامل

Microdosimetry for boron neutron capture therapy.

Preclinical studies for boron neutron capture therapy (BNCT) using epithermal neutrons are ongoing at several laboratories. The absorbed dose in tumor cells is a function of the thermal neutron flux at depth, the microscopic boron concentration, and the size of the cell. Dosimetry is therefore complicated by the admixture of thermal, epithermal, and fast neutrons, plus gamma rays, and the array...

متن کامل

An investigation into the potential applicability of gel dosimeters for dosimetry in boron neutron capture therapy

Background: The aim of this work was to establish how well gel dosimeters performed, as substitutes for brain tissue compared with standard phantom materials such as water, polymethyl-methacrylate (or PMMA), A150 plastic and TE- liquid phantom material for dosimetry of neutron beams in boron neutron capture therapy. Materials and Methods: Thermal neutron fluence, photon dose and epithermal neu...

متن کامل

Finnish dosimetric practice for epithermal neutron beam dosimetry in boron neutron capture therapy

Boron neutron capture therapy (BNCT) is a form of chemically targeted radiotherapy that utilises the high neutron capture cross-section of boron-10 isotope to achieve a preferential dose increase in the tumour. The BNCT dosimetry poses a special challenge as the radiation dose absorbed by the irradiated tissues consists of several dose components with different relative biological effectiveness...

متن کامل

Investigation the potential of Boron neutron capture therapy (BNCT) to treat the lung cancer

Introduction: Boron neutron capture therapy (BNCT) is recommended to treat the glioblastoma tumor. It is well known that neuron beams are more effective treatment than photon beams to treat hypoxia tumors due to interaction of neutron with nucleus and production of heavy particles such as 7Li and alpha particle. In this study to evaluate the suitability of BNCT for treating of ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 25 11  شماره 

صفحات  -

تاریخ انتشار 1998