Bone-composition imaging using coherent-scatter computed tomography: assessing bone health beyond bone mineral density.
نویسندگان
چکیده
Quantitative analysis of bone composition is necessary for the accurate diagnosis and monitoring of metabolic bone diseases. Accurate assessment of the bone mineralization state is the first requirement for a comprehensive analysis. In diagnostic imaging, x-ray coherent scatter depends upon the molecular structure of tissues. Coherent-scatter computed tomography (CSCT) exploits this feature to identify tissue types in composite biological specimens. We have used CSCT to map the distributions of tissues relevant to bone disease (fat, soft tissue, collagen, and mineral) within bone-tissue phantoms and an excised cadaveric bone sample. Using a purpose-built scanner, we have measured hydroxyapatite (bone mineral) concentrations based on coherent-scatter patterns from a series of samples with varying hydroxyapatite content. The measured scatter intensity is proportional to mineral density in true g/cm3. Repeated measurements of the hydroxyapatite concentration in each sample were within, at most, 2% of each other, revealing an excellent precision in determining hydroxyapatite concentration. All measurements were also found to be accurate to within 3% of the known values. Phantoms simulating normal, over-, and under-mineralized bone were created by mixing known masses of pure collagen and hydroxyapatite. An analysis of the composite scatter patterns gave the density of each material. For each composite, the densities were within 2% of the known values. Collagen and hydroxyapatite concentrations were also examined in a bone-mimicking phantom, incorporating other bone constituents (fat, soft tissue). Tomographic maps of the coherent-scatter properties of each specimen were reconstructed, from which material-specific images were generated. Each tissue was clearly distinguished and the collagen-mineral ratio determined from this phantom was also within 2% of the known value. Existing bone analysis techniques cannot determine the collagen-mineral ratio in intact specimens. Finally, to demonstrate the in situ potential of this technique, the mineralization state of an excised normal cadaveric radius was examined. The average collagen-mineral ratio of the cortical bone derived from material-specific images of the radius was 0.53+/-0.04, which is in agreement with the expected value of 0.55 for healthy bones.
منابع مشابه
Material-specific analysis using coherent-scatter imaging.
Coherent-scatter computed tomography (CSCT) is a novel imaging method we are developing to produce cross-sectional images based on the low-angle (<10 degrees) scatter properties of tissue. At diagnostic energies, this scatter is primarily coherent with properties dependent upon the molecular structure of the scatterer. This facilitates the production of material-specific maps of each component ...
متن کاملAssessment of bone fragility with clinical imaging modalities.
INTRODUCTION Osteoporotic fractures are a vital public health concern and have created a great economic burden to our society. Therefore, early diagnosis of patients with high risk of osteoporotic fractures is essential. The current gold standard for assessment of fracture risk is the measurement of bone mineral density using dual-energy X-ray absorptiometry. However, such techniques are not ve...
متن کاملBone strength and management of postmenopausal fracture risk with antiresorptive therapies: considerations for women’s health practice
Bone strength - and, hence, fracture risk - reflects the structural and material properties of the skeleton, which changes with bone turnover during aging and following effective pharmacotherapy. A variety of powerful new techniques (quantitative computed tomography, as well as peripheral quantitative computed tomography and high-resolution peripheral quantitative computed tomography) provide p...
متن کاملReconstructionof Human Mandibular Continuity Defects with Allogenic Scaffold and Autologousmarrow Mesenchymal Stem Cells
Background Mandibular continuity defects occur after tumor resection, maxillofacial injury, or osteomyelitis. Despite the current availability of a plethora of treatment modalities, bone substitutes, and various clinical adjuncts, an exact reconstructive recapitulation of large bony defects continues to be beyond reach. In this clinical pilot study, we report a novel method for reconstruction ...
متن کاملCorrelation Between Tissue Densities in Computed Tomography and Three Different Cone-Beam Computed Tomography Units (In Vitro)
Background and Aim: Bone density is of great assistance in the selection of the proper implant site. The present study aimed to assess the correlation between tissue densities in computed tomography (CT) and three different cone-beam computed tomography (CBCT) units. Materials and Methods: In this descriptive study, a radiographic phantom consisting of a transparent polymethyl methacrylate (PM...