Comparison of five radioimmunoassays and enzyme bioassay for measurement of digoxin in blood.

نویسندگان

  • A A MacKinney
  • G H Burnett
  • R L Conklin
  • G W Wasson
چکیده

Digoxin-containing sera from 86 patients were analyzed for this drug by bioassay and radioimmunoassay. Each serum was analyzed in duplicate by six procedures: inhibition of Na+-K+-dependent ATPase and five "kit" radioimmunoassays from four different commercial sources. Mean values for two of the radioimmunoassays differed significantly from those for the bioassay. One radioimmunoassay mean value was significantly different from the other five mean values. We conclude that normal values for digoxin radioimmunoassay should be determined for each kit, and should not be adopted from published data.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of Honey on CYP3A4 Enzyme and P-Glycoprotein Activity in Healthy Human Volunteers

The activity of cytochrome p450 isozyme 3A4 (CYP3A4) enzyme and P-glycoprotein (P-gp) is modulated by grapefruit juice and herbal drugs. CYP3A4 is the major phase I drug metabolizing enzyme and P-gp is an ATP-dependent drug efflux pump that regulates the intestinal absorption of orally administered drugs. Honey is commonly consumed as a dietary supplement. However, its influence on human CY...

متن کامل

Inspection of Digoxin level in cardiac heart failure patients refer to Taemin Ejtemai and Shohadaye Ashaer clinic

Background: Heart failure is one of the common cardiovascular diseases and digoxin has the positive effect on it. But unfortunately in spite of its positive effect on heart failure therapeutic and toxic level of it in patients is different and close to each other. The aim of this research was determination of digoxin serum level in failur heart patients. Materials and methods: In this cros...

متن کامل

Radioimmunoassay and enzyme immunoassay compared for determination of digoxin.

Patients' sera were analyzed for digoxin by using two different radioimmunoassays and an enzyme immunoassay. Quantitative results obtained by enzyme immunoassay (I) were compared to results obtained on aliquots of the same sample by the radioimmunoassays (II and III). The correlation coefficients were: I vs. II 0.90, n=108; I vs. III 0.94, n=102; and II vs. III 0.95, n=158. Day-to-day precision...

متن کامل

Determination of the Effects of Digoxin on the Right Ventricular Function in Patients Undergoing Pneumonectomy

Introduction: Pneumonectomy is the standard treatment of lung cancer, even though patients should undergo several evaluations before surgery; deterioration of cardiopulmonary function after pulmonary resection is inevitable. We have evaluated the effects of digoxin on the improvement of right ventricular function and prevention of probable complications after lung resection surgery. Materials a...

متن کامل

ENZYME LEVELS IN THE SERA AND ERYTHROCYTES OF PATIENTS WITH VISCERAL LEISH MANIASIS

Twenty five blood samples from 6-month to 5-year old children with visceral leishmaniasis were analysed for various enzymes. Aspartate transaminase (AST), alanine transaminase (ALT), lactate dehydrogenase (LD) and its various isoenzymes were estimated in the serum, while glucose 6-phosphate dehydrogenase (G6PD) was measured in red blood cells. For comparison, blood samples from healthy chi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 21 7  شماره 

صفحات  -