Errors introduced by dose scaling for relative dosimetry
نویسندگان
چکیده
Some dosimeters require a relationship between detector signal and delivered dose. The relationship (characteristic curve or calibration equation) usually depends on the environment under which the dosimeters are manufactured or stored. To compensate for the difference in radiation response among different batches of dosimeters, the measured dose can be scaled by normalizing the measured dose to a specific dose. Such a procedure, often called "relative dosimetry", allows us to skip the time-consuming production of a calibration curve for each irradiation. In this study, the magnitudes of errors due to the dose scaling procedure were evaluated by using the characteristic curves of BANG3 polymer gel dosimeter, radiographic EDR2 films, and GAFCHROMIC EBT2 films. Several sets of calibration data were obtained for each type of dosimeters, and a calibration equation of one set of data was used to estimate doses of the other dosimeters from different batches. The scaled doses were then compared with expected doses, which were obtained by using the true calibration equation specific to each batch. In general, the magnitude of errors increased with increasing deviation of the dose scaling factor from unity. Also, the errors strongly depended on the difference in the shape of the true and reference calibration curves. For example, for the BANG3 polymer gel, of which the characteristic curve can be approximated with a linear equation, the error for a batch requiring a dose scaling factor of 0.87 was larger than the errors for other batches requiring smaller magnitudes of dose scaling, or scaling factors of 0.93 or 1.02. The characteristic curves of EDR2 and EBT2 films required nonlinear equations. With those dosimeters, errors larger than 5% were commonly observed in the dose ranges of below 50% and above 150% of the normalization dose. In conclusion, the dose scaling for relative dosimetry introduces large errors in the measured doses when a large dose scaling is applied, and this procedure should be applied with special care.
منابع مشابه
Experimental evaluation of midline dose calculation methods in In vivo dosimetry using anatomic thorax phantom
Background: In vivo dosimetry is a method for estimation of overall error in the delivered dose to the patients at the end of radiotherapy process. In this research, two methods for target dose calculation were evaluated on midline and central axis of photon beams in in vivo dosimetry of thorax fields. Materials and Methods: Entrance and exit doses for anterior and lateral fields of thorax were...
متن کاملDeveloping a Verification and Training Phantom for Gynecological Brachytherapy System
Introduction Dosimetric accuracy is a major issue in the quality assurance (QA) program for treatment planning systems (TPS). An important contribution to this process has been a proper dosimetry method to guarantee the accuracy of delivered dose to the tumor. In brachytherapy (BT) of gynecological (Gyn) cancer it is usual to insert a combination of tandem and ovoid applicators with a complicat...
متن کاملمروری بر انواع دزیمترهای سه بعدی
Modern radiotherapy techniques try to deliver the radiation dose by steep gradient. Protocols for these complex treatment delivery techniques are important and extremely sensitive to errors in treatment delivery. To minimize such errors, 3D dosimeters were developed as a relative method for improving dose monitoring and delivery. The aim of this study is to analyze the capability and drawbacks ...
متن کاملبررسی صحیح بودن مقدار اشعه داده شده در پرتودرمانی سرطان پستان با ترمولومینسانس دوزیمتری
In vivo dosimetry seems an appropriate method for quality assurance in radiotherapy with CO-60, which can be considered as a routine program in radiotherapy centers. In this study, entrance and exit dose in tangential fields were measured in phantom for 30 breast cancer patients by using TLD in radiotherapy section of Imam Khomeini Hospital. TLD crystals were calibrated against SSDL (Secondary...
متن کاملUncertainty Analysis in MRI-based Polymer Gel Dosimetry
Background: Polymer gel dosimeters combined with magnetic resonance imaging (MRI) can be used for dose verification of advanced radiation therapy techniques. However, the uncertainty of dose map measured by gel dosimeter should be known. The purpose of this study is to investigate the uncertainty related to calibration curve and MRI protocol for MAGIC (Methacrylic and Ascorbic acid in Gelatin I...
متن کامل