Initial experience in the extraction of chronically implanted pacemaker leads using the Excimer laser sheath.

نویسندگان

  • T Levy
  • S Walker
  • V Paul
چکیده

OBJECTIVE To assess the safety and efficiency of the Excimer laser sheath in extracting chronically implanted pacemaker leads. PATIENTS Eight patients were studied (one female, mean age 62 years, range 34 to 77) with 17 pacemaker leads (five atrial, 10 ventricular, two implantable defibrillator). The mean implantation time was 65 months (range 23 to 188). The indications for lead extraction were chronic infection (7), superior vena cava obstruction (4), lead malfunction (4), and pain (2). METHODS A prospective analysis of the use of the Excimer laser sheath in extracting chronically implanted pacemaker leads. Laser sheath extraction was undertaken if conventional extraction techniques with simple traction or traction with a locking stylet had failed. If laser sheath extraction was unsuccessful, basket retrieval of the lead from the groin was performed. RESULTS Complete lead removal was achieved in 16 leads (94%). In one case the electrode tip was left behind without complication. Extraction was achieved with the laser sheath alone in 16 leads. Basket retrieval was required in one case after laser failure. There were no complications. CONCLUSIONS The Excimer laser sheath appears to be an effective and safe technique for extracting chronically implanted pacemaker leads. It can be used in combination with the currently available techniques for successful lead extraction.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pacemaker lead extraction with the laser sheath: results of the pacing lead extraction with the excimer sheath (PLEXES) trial.

OBJECTIVES The purpose of this study was to evaluate the safety and effectiveness of pacemaker lead extraction with the excimer sheath in comparison to nonlaser lead extraction. BACKGROUND Fibrotic attachments that develop between chronically implanted pacemaker leads and to the venous, valvular and cardiac structures are the major obstacles to safe and consistent lead extraction. Locking sty...

متن کامل

Determinants of procedural outcome of chronically implanted pacemaker and defibrillator leads using the Excimer laser sheath.

The extraction of chronic pacemaker and internal cardioverter defibrillator leads is performed for a number of reasons including chronic infection, lead dysfunction, and venous obstruction. Traditionally leads were removed by traction which is associated with a high incidence of failure and serious complications. This has led to the introduction of laser lead extraction (LLE) which is a safe an...

متن کامل

[Initial experience with excimer laser-assisted pacemaker and defibrillator lead extraction].

Explantation of chronically implanted pacemaker or defibrillator leads can be technically demanding. We present the results of the first 44 leads extracted with excimer laser sheaths in 25 patients with a median lead implantation time of 7 years (range: 2 to 20 years). Indications for extraction were lead fracture in 17 patients, pocket infection in 4, septicemia in 2, and lead endocarditis in ...

متن کامل

Transvenous extraction of a 26-year-old Accufix atrial lead using TightRail rotating dilator sheath.

With the increasing number of implanted pacemakers and implantable cardioverter defibrillators, removal is required more frequently. Presently described is the transvenous extraction of a 26-year-old Accufix atrial lead using a mechanical dilator sheath. A 50-year-old male patient was admitted to the clinic with a pacemaker pocket infection. The atrial lead was an Accufix Bipolar J-Atrial activ...

متن کامل

An Implantable Left Atrial Pressure Sensor Lead Designed for Percutaneous Extraction Using Standard Techniques

BACKGROUND An implantable left atrial pressure (LAP) monitoring system for guiding the management of patients with advanced heart failure has the potential to require extraction, particularly in the setting of infection. The LAP sensor lead was designed to be suitable for ease of percutaneous extraction using standard techniques for extracting pacemaker and defibrillator leads. The clinical exp...

متن کامل