TrueBeam Low Dose Rate Investigation for Pulsed Reduced Dose Rate IMRT

نویسنده

  • Mark Geurts
چکیده

The capability of the TrueBeam treatment system to deliver step and shoot IMRT plans at low dose rates was evaluated. Beam characteristics during low dose rates (5 to 100 MU/min) were evaluated for consistency using a planar ion chamber array. MU constancy, linearity, and beam profile symmetry were all found to be equivalent within 0.5%. The response of the Scandi Dos Delta 4 system was also evaluated at low dose rates of using static open beams compared to ion chamber measurements, and step and shoot IMRT plans comparing 5 20 MU/min and 100 MU/min dose rates, with a maximum observed absolute dose difference of 0.8% and equivalence margin of 0.2%. The Gamma Index and measurement reproducibility were also found to be equivalent.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The effect of high energy photon beam in dose- volume parameters and integral dose of normal structures in Head and neck IMRT

Introduction: The aim of this study is to investigate the effect of partially used high photon energies on dosimetric parameters and integral doses of nasopharynx IMRT plans.   Materials and Methods: Two 7 coplanar IMRT plan were created for eleven patients (6 MV only IMRT plan in both phase1 and phase2 and 6 MV phase1-mixed energies (6 MV and 15 MV) phase2...

متن کامل

Optimization strategies for pulsed low‐dose‐rate IMRT of recurrent lung and head and neck cancers

Pulsed low-dose-rate radiotherapy (PLDR) has been proven to be a valid method of reirradiation. Previous studies of recurrent cancer radiotherapy were mainly based on conventional 3D CRT and VMAT delivery techniques. There are difficulties in IMRT planning using existing commercial treatment planning systems (TPS) to meet the PLDR protocol. This work focuses on PLDR using ten-field IMRT and a c...

متن کامل

The influence of respiratory motion on dose distribution of 3D-CRT and IMRT- A simulation study

Background: 3DCRT (three-dimensional conformal radiotherapy) and IMRT (intensity-modulated radiotherapy) has provided us with tools to delineate the radiation dose distribution of tumor targets. However, the precision of radiation can be compromised by respiratory motion, which usually limits the geometric and dosimetric accuracy of radiotherapy. The purpose of this study is to evaluate the imp...

متن کامل

Use of reduced dose rate when treating moving tumors using dynamic IMRT

The purpose was to evaluate the effect of dose rate on discrepancies between expected and delivered dose caused by the interplay effect. Fifteen separate dynamic IMRT plans and five hybrid IMRT plans were created for five patients (three IMRT plans and one hybrid IMRT plan per patient). The impact of motion on the delivered dose was evaluated experimentally for each treatment field for differen...

متن کامل

IMRT and RapidArc commissioning of a TrueBeam linear accelerator using TG‐119 protocol cases

The purpose of this study is to evaluate the overall accuracy of intensity-modulated radiation therapy (IMRT) and RapidArc delivery using both flattening filter (FF) and flattening filter-free (FFF) modalities based on test cases developed by AAPM Task Group 119. Institutional confidence limits (CLs) were established as the baseline for patient specific treatment plan quality assurance (QA). Th...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2017