On flattening filter‐free portal dosimetry
نویسندگان
چکیده
Varian introduced (in 2010) the option of removing the flattening filter (FF) in their C-Arm linacs for intensity-modulated treatments. This mode, called flattening filter-free (FFF), offers the advantage of a greater dose rate. Varian's "Portal Dosimetry" is an electronic portal imager device (EPID)-based tool for IMRT verification. This tool lacks the capability of verifying flattening filter-free (FFF) modes due to saturation and lack of an image prediction algorithm. (Note: the latest versions of this software and EPID correct these issues.) The objective of the present study is to research the feasibility of said verifications (with the older versions of the software and EPID). By placing the EPID at a greater distance, the images can be acquired without saturation, yielding a linearity similar to the flattened mode. For the image prediction, a method was optimized based on the clinically used algorithm (analytical anisotropic algorithm (AAA)) over a homogeneous phantom. The depth inside the phantom and its electronic density were tailored. An application was developed to allow the conversion of a dose plane (in DICOM format) to Varian's custom format for Portal Dosimetry. The proposed method was used for the verification of test and clinical fields for the three qualities used in our institution for IMRT: 6X, 6FFF and 10FFF. The method developed yielded a positive verification (more than 95% of the points pass a 2%/2 mm gamma) for both the clinical and test fields. This method was also capable of "predicting" static and wedged fields. A workflow for the verification of FFF fields was developed. This method relies on the clinical algorithm used for dose calculation and is able to verify the FFF modes, as well as being useful for machine quality assurance. The procedure described does not require new hardware. This method could be used as a verification of Varian's Portal Dose Image Prediction.
منابع مشابه
Enhancement of the Dose on 12 MV Linac with Free Flattening Filter Mode
Purpose: In the last years, some studies investigated dosimetric benefits of a free flattening filter for the photon mode in the radiotherapy field. This study aims to provide a theoretical study published and analysis of basic dosimetric properties for a Saturne 43 Linac head to implement free flattening filter beams clinically. Material and Methods: Thi...
متن کاملPractical approach for pretreatment verification of IMRT with flattening filter‐free (FFF) beams using Varian portal dosimetry
Patient-specific pretreatment verification of intensity-modulated radiation therapy (IMRT) or volumetric-modulated arc therapy (VMAT) is strongly recommended for all patients in order to detect any potential errors in treatment planning process and machine deliverability, and is thus performed routinely in many clinics. Portal dosimetry is an effective method for this purpose because of its pro...
متن کاملFeasibility of portal dosimetry for flattening filter‐free radiotherapy
The feasibility of using portal dosimetry (PD) to verify 6 MV flattening filter-free (FFF) IMRT treatments was investigated. An Elekta Synergy linear accelerator with an Agility collimator capable of delivering FFF beams and a standard iViewGT amorphous silicon (aSi) EPID panel (RID 1640 AL5P) at a fixed SSD of 160 cm were used. Dose rates for FFF beams are up to four times higher than for conv...
متن کاملComparison between the TRS-398 code of practice and the TG-51 dosimetry protocol for flattening filter free beams.
Dosimetry protocols for external beam radiotherapy currently in use, such as the IAEA TRS-398 and AAPM TG-51, were written for conventional linear accelerators. In these accelerators, a flattening filter is used to produce a beam which is uniform at water depths where the ionization chamber is used to measure the absorbed dose. Recently, clinical linacs have been implemented without the flatten...
متن کاملDosimetric influence of Flattening Filter (FF) and Flattening Filter Free (FFF) 6 and 10 MV photon beams on Volumetric Modulated Arc Therapy (VMAT) planning in case of prostate carcinoma
Background: In the treatment of prostate cancer, radiotherapy is the potential to increase second primary cancers such as bladder and rectal cancers. The reasons for this potential are more monitor units (MUs), therefore a larger total body dose because of leakage radiation, a bigger volume of normal tissue is exposed to lower radiation doses. This study was designed to compare the integral dos...
متن کامل