Performance measurement of the microPET focus 120 scanner.
نویسندگان
چکیده
UNLABELLED The microPET Focus 120 scanner is a third-generation animal PET scanner dedicated to rodent imaging. Here, we report the results of scanner performance testing. METHODS A (68)Ge point source was used to measure energy resolution, which was determined for each crystal and averaged. Spatial resolution was measured using a (22)Na point source with a nominal size of 0.25 mm at the system center and various off-center positions. Absolute sensitivity without attenuation was determined by extrapolating the data measured using an (18)F line source and multiple layers of absorbers. Scatter fraction and counting rate performance were measured using 2 different cylindric phantoms simulating rat and mouse bodies. Sensitivity, scatter fraction, and noise equivalent counting rate (NECR) experiments were repeated under 4 different conditions (energy window, 250 approximately 750 keV or 350 approximately 650 keV; coincidence window, 6 or 10 ns). A performance phantom with hot-rod inserts of various sizes was scanned, and several animal studies were also performed. RESULTS Energy resolution at a 511-keV photopeak was 18.3% on average. Radial, tangential, and axial resolution of images reconstructed with the Fourier rebinning (FORE) and filtered backprojection (FBP) algorithms were 1.18 (radial), 1.13 (tangential), and 1.45 mm full width at half maximum (FWHM) (axial) at center and 2.35 (radial), 1.66 (tangential), and 2.00 mm FWHM (axial) at a radial offset of 2 cm. Absolute sensitivities at transaxial and axial centers were 7.0% (250 approximately 750 keV, 10 ns), 6.7% (250 approximately 750 keV, 6 ns), 4.0% (350 approximately 650 keV, 10 ns), and 3.8% (350 approximately 650 keV, 6 ns). Scatter fractions were 15.9% (mouse phantom) and 35.0% (rat phantom) for 250 approximately 750 keV and 6 ns. Peak NECR was 869 kcps at 3,242 kBq/mL (mouse phantom) and 228 kcps at 290 kBq/mL (rat phantom) at 250 approximately 750 keV and 6 ns. Hot-rod inserts of 1.6-mm diameter were clearly identified, and animal studies illustrated the feasibility of this system for studies of whole rodents and mid-sized animal brains. CONCLUSION The results of this independent field test showed the improved physical characteristics of the F120 scanner over the previous microPET series systems. This system will be useful for imaging studies on small rodents and brains of larger animals.
منابع مشابه
Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging.
UNLABELLED The microPET Focus is the latest generation microPET system dedicated to high-resolution animal imaging and incorporates several changes to enhance its performance. This study evaluated the basic performance of the scanner and compared it with the Primate (P4) and Rodent (R4) models. METHODS The system consists of 168 lutetium oxyorthosilicate (LSO) detectors arranged in 4 contiguo...
متن کاملMeasurement of organ dose in abdomen-pelvis CT exam as a function of mA, KV and scanner type by Monte Carlo method .
Background: CT is a diagnostic imaging modality giving higher patient dose in comparison with other radiological procedures, so the calculation of organ dose in CT exams is very important. While methods to calculate the effective dose have been established (ICRP 26 and ICRP 60), they depend heavily on the ability to estimate the dose to radiosensitive organs from the CT procedure. However, dete...
متن کاملOptimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging.
MicroPET II is a newly developed PET (positron emission tomography) scanner designed for high-resolution imaging of small animals. It consists of 17,640 LSO crystals each measuring 0.975 x 0.975 x 12.5 mm3, which are arranged in 42 contiguous rings, with 420 crystals per ring. The scanner has an axial field of view (FOV) of 4.9 cm and a transaxial FOV of 8.5 cm. The purpose of this study was to...
متن کاملPerformance evaluation of microPET: a high-resolution lutetium oxyorthosilicate PET scanner for animal imaging.
UNLABELLED A new dedicated PET scanner, microPET, was designed and developed at the University of California, Los Angeles, for imaging small laboratory animals. The goal was to provide a compact system with superior spatial resolution at a fraction of the cost of a clinical PET scanner. METHODS The system uses fiberoptic readout of individually cut lutetium oxyorthosilicate (LSO) crystals to ...
متن کاملPerformance evaluation of an Inveon PET preclinical scanner.
We evaluated the performance of an Inveon preclinical PET scanner (Siemens Medical Solutions), the latest MicroPET system. Spatial resolution was measured with a glass capillary tube (0.26 mm inside diameter, 0.29 mm wall thickness) filled with (18)F solution. Transaxial and axial resolutions were measured with the source placed parallel and perpendicular to the axis of the scanner. The sensiti...