Loss of sagittal alignment and balance in adult spinal deformity can cause severe pain, disability and progressive neurological deficit. When conservative treatment has failed, spinal fusion using rigid instrumentation is currently the salvage treatment to stop further curve progression. However, fusion surgery is associated with high revision rates due to instrumentation failure and proximal junctional failure, especially if patients also suffer from osteoporosis. To address these drawbacks, a less rigid rod construct is proposed, which is hypothesized to provide a more gradual transition of force and load distribution over spinal segments in comparison to stiff titanium rods. In this study, the effect of variation in rod stiffness on the intradiscal pressure (IDP) of fixed spinal segments during flexion-compression loading was assessed. An ex vivo multisegment (porcine) flexion-compression spine test comparing rigid titanium rods with more flexible polycarbonate-urethane (PCU) rods was used. An increase in peak IDP was found for both the titanium and PCU instrumentation groups as compared to the uninstrumented controls. The peak IDP for the spines instrumented with the PCU rods was significantly lower in comparison to the titanium instrumentation group. These results demonstrated the differences in mechanical load transfer characteristics between PCU and titanium rod constructs when subjected to flexion-compression loading. The concept of stabilization with a less rigid rod may be an alternative to fusion with rigid instrumentation, with the aim of decreasing mechanical stress on the instrumented segments and the possible benefit of a decrease in the incidence of screw pullout.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Posterior spinal fixation with pedicle screws and rods system in thoracolumbar spinal fractures.

OBJECTIVE To determine the effectiveness of posterior spinal fixation in maintaining the stability of spine and to determine mortality in 6 months after posterior spinal fixation. STUDY DESIGN Quasi experimental study. PLACE AND DURATION OF STUDY Department of Orthopaedics, Jinnah Postgraduate Medical Centre, Karachi and Department of Orthopaedics, Postgraduate Medical Institute, Hayatabad ...

متن کامل

Thoracolumbar junction injuries and their management with pedicle screws.

BACKGROUND To evaluate the use of pedicle screw fixation in earthquake injured thoracolumbar spine. METHODS Nineteen patients with posttraumatic instability of lower thoracic or upper lumbar spine were included in the study. White and Panjabi criteria was used to assess spinal instability. All patients underwent open reduction and internal fixation by posterior approach. Pedicles were localiz...

متن کامل

Assessment of the suitability of biodegradable rods for use in posterior lumbar fusion: An in-vitro biomechanical evaluation and finite element analysis

Interbody fusion with posterior instrumentation is a common method for treating lumbar degenerative disc diseases. However, the high rigidity of the fusion construct may produce abnormal stresses at the adjacent segment and lead to adjacent segment degeneration (ASD). As such, biodegradable implants are becoming more popular for use in orthopaedic surgery. These implants offer sufficient stabil...

متن کامل

Clinical and Radiological Comparison of Semirigid (WavefleX) and Rigid System for the Lumbar Spine

OBJECTIVE Spinal fusion operation is an effective treatment in the spinal pathology, but it could change the physiological distribution of load at the instrumented and adjacent segments. This retrospective study compared the radiological and clinical outcomes of patients undergoing lumbar fusion with semirigid rods versus rigid rods system. METHODS Using transpedicular fixation and posterior ...

متن کامل

Does Semi-Rigid Instrumentation Using Both Flexion and Extension Dampening Spacers Truly Provide an Intermediate Level of Stabilization?

Conventional posterior dynamic stabilization devices demonstrated a tendency towards highly rigid stabilization approximating that of titanium rods in flexion. In extension, they excessively offload the index segment, making the device as the sole load-bearing structure, with concerns of device failure. The goal of this study was to compare the kinematics and intradiscal pressure of monosegment...