Canted seating of the Stryker Modular Dual Mobility liner within a Trident hemispherical acetabular shell

نویسندگان

  • Scott M. Eskildsen
  • Erik C. Olsson
  • Daniel J. Del Gaizo
چکیده

A 75-year-old woman who suffered a left femoral neck fracture underwent a left total hip arthroplasty using a Stryker Trident (Kalamazoo, MI) hemispherical acetabular shell and Modular Dual Mobility (MDM) metal liner. Post-operative radiographs demonstrated canted seating of the liner. The patient was taken immediately back to the operating room where the acetabular liner appeared well seated superiorly but was in a canted position inferiorly. Removal and replacement was performed and post-operative radiographs demonstrated complete seating. Subsequent follow up at 6 months demonstrated good clinical function with no adverse radiographic findings. Canted seating is a potential complication of the MDM metal liner. Providers should be aware of potential incomplete seating inferiorly despite the superior portion of the liner being well seated.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Incomplete seating of the liner with the Trident acetabular system: a cause for concern?

We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteropos...

متن کامل

Dual poly liner mobility optimizes wear and stability in THA: affirms.

A dual-mobility acetabular component consists of a large, fixed, porous-coated acetabular component and a bipolar femoral component. These components are often called tripolar components. This configuration provides a stable, well-fixed implant platform against bone and 2 articular interfaces, a large polyethylene surface directly apposed to a highly polished metal implant, and a standard-sized...

متن کامل

Cementation of a dual-mobility acetabular component into a well-fixed metal shell during revision total hip arthroplasty: a biomechanical validation.

Cementation of polyethylene (PE) liners into well-fixed metal shells has become a popular option during revision total hip arthroplasty (THA) particularly for older and frail patients. Although dramatic results were reported with dual-mobility acetabular components to manage hip instability during revision THA, no study evaluated the fixation strength of the cementation of dual-mobility compone...

متن کامل

Early Intraprosthetic Dislocation of A Dual Mobility Acetabular Construct

Dislocations represent one of the more common postoperative complications seen after total hip arthroplasty (THA) with estimates ranging from 0.2 to 7% in primary cases and as high as 25% in revision surgeries [1]. In the early 1970s, Giles Bousquet developed a dual mobility (DM) acetabular cup with the goal of reducing dislocation rates. The dual mobility (DM) design consists of two articulati...

متن کامل

Do thin acetabular shells increase the disassociation risk of ceramic liners?

This study suggests that contemporary modular acetabular shell designs employing ceramic liners deform during implantation where acetabular bed under-reaming is employed. The liner retention strength did not degrade by comparison to non-deformed shells using ASTM static push-out methods. A larger question, however, is the influence of this deformation on liner retention strength over time when ...

متن کامل