Shortening of atrioventricular delay at increased atrial paced heart rates improves diastolic filling and functional class in patients with biventricular pacing
نویسندگان
چکیده
BACKGROUND Use of rate adaptive atrioventricular (AV) delay remains controversial in patients with biventricular (Biv) pacing. We hypothesized that a shortened AV delay would provide optimal diastolic filling by allowing separation of early and late diastolic filling at increased heart rate (HR) in these patients. METHODS 34 patients (75 ± 11 yrs, 24 M, LVEF 34 ± 12%) with Biv and atrial pacing had optimal AV delay determined at baseline HR by Doppler echocardiography. Atrial pacing rate was then increased in 10 bpm increments to a maximum of 90 bpm. At each atrial pacing HR, optimal AV delay was determined by changing AV delay until best E and A wave separation was seen on mitral inflow pulsed wave (PW) Doppler (defined as increased atrial duration from baseline or prior pacemaker setting with minimal atrial truncation). Left ventricular (LV) systolic ejection time and velocity time integral (VTI) at fixed and optimal AV delay was also tested in 13 patients. Rate adaptive AV delay was then programmed according to the optimal AV delay at the highest HR tested and patients were followed for 1 month to assess change in NYHA class and Quality of Life Score as assessed by Minnesota Living with Heart Failure Questionnaire. RESULTS 81 AV delays were evaluated at different atrial pacing rates. Optimal AV delay decreased as atrial paced HR increased (201 ms at 60 bpm, 187 ms at 70 bpm, 146 ms at 80 bpm and 123 ms at 90 bpm (ANOVA F-statistic = 15, p = 0.0010). Diastolic filling time (P < 0.001 vs. fixed AV delay), mitral inflow VTI (p < 0.05 vs fixed AV delay) and systolic ejection time (p < 0.02 vs. fixed AV delay) improved by 14%, 5% and 4% respectively at optimal versus fixed AV delay at the same HR. NYHA improved from 2.6 ± 0.7 at baseline to 1.7 ± 0.8 (p < 0.01) 1 month post optimization. Physical component of Quality of Life Score improved from 32 ± 17 at baseline to 25 ± 12 (p < 0.05) at follow up. CONCLUSIONS Increased heart rate by atrial pacing in patients with Biv pacing causes compromise in diastolic filling time which can be improved by AV delay shortening. Aggressive AV delay shortening was required at heart rates in physiologic range to achieve optimal diastolic filling and was associated with an increase in LV ejection time during optimization. Functional class improved at 1 month post optimization using aggressive AV delay shortening algorithm derived from echo-guidance at the time of Biv pacemaker optimization.
منابع مشابه
Non-invasive Determination of the Optimized Atrioventricular Delay in Patients with Implanted Biventricular Pacing Devices
BACKGROUND Biventricular (BiV) is extensively used in the treatment of congestive heart failure but so far no recommendations for optimized programming of atrioventricular-delay (AVD) settings have been proposed. Can AVD optimization be performed using a simple formula based on non-invasive doppler-echocardiography? METHODS 25 patients (ejection fraction 30+/-8%) received BiV ICDs. Doppler-ec...
متن کاملLong-term follow-up of atrioventricular delay optimization in patients with biventricular pacing.
BACKGROUND Atrioventricular (AV) delay optimization may be important in patients with biventricular pacing and the optimal AV delay can be predicted using Doppler echocardiography and the formula: optimal AV delay = AV delay-the interval between the end of A wave and complete closure of the mitral valve when the AV delay is set at slightly prolonged AV delay. METHODS AND RESULTS In the presen...
متن کاملEchocardiographic Demonstration of the Effect of Varying Paced A-V Intervals on Ventricular Filling Pattern
Impaired relaxation of the left ventricle (LV) (diastolic dysfunction) invariably is associated with LV systolic dysfunction.1 Atrial contribution to ventricular filling (second filling phase of diastole) is pivotal in the presence of impaired relaxation where early filling is reduced. Cardiac Resynchronization Therapy (CRT) has been shown to improve diastolic function in patients with heart fa...
متن کاملImprovement in coronary blood flow velocity with acute biventricular pacing is predominantly due to an increase in a diastolic backward-travelling decompression (suction) wave.
BACKGROUND Normal coronary blood flow is principally determined by a backward-traveling decompression (suction) wave in diastole. Dyssynchronous chronic heart failure may attenuate suction, because regional relaxation and contraction overlap in timing. We hypothesized that biventricular pacing, by restoring left ventricular (LV) synchronization and improving LV relaxation, might increase this s...
متن کاملCardiac resynchronization in patients with congestive heart failure and chronic atrial fibrillation: effect of upgrading to biventricular pacing after chronic right ventricular pacing.
OBJECTIVES This study assessed the effects of biventricular pacing (BVP) on ventricular function, functional status, quality of life and hospitalization in patients with congestive heart failure (CHF), prior atrioventricular (AV) junction ablation and right ventricular (RV) pacing performed for chronic atrial fibrillation (AF). BACKGROUND Although the benefit of BVP in CHF should theoreticall...