Repeatability and reproducibility of upper eyelid measurements.
نویسندگان
چکیده
AIM The aim of this study was to assess the repeatability and reproducibility by physicians of upper lid measurements and to investigate the influence of clinical experience on the learning curve effect. METHODS Both eyes of 22 outpatients were assessed for three basic measures of ptosis: marginal reflex distance (MRD) for upper and lower lids, upper lid skin crease (SC), and levator function (LF). Patients with variable eyelid positions were excluded. The patients were measured twice by a consultant and once by each of a clinical fellow, a specialist registrar, and a senior house officer in random order. Each observer was masked to their colleagues' results and followed a standard measurement protocol. Data were analysed using Bland-Altman plots. RESULTS Consultant repeatability was high and consistent, the median difference between measures being 0 for each of the four parameters. Clinically acceptable reproducibility was shown in all measurements for even the least experienced physician and was particularly consistent for extreme observations. There was evidence of a learning curve effect. CONCLUSIONS These results suggest that interobserver and intraobserver variability in assessment of upper lid ptosis using a standard measurement protocol is low and clinically acceptable when the technique of assessment is standardised.
منابع مشابه
Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents
Purpose. To assess the repeatability and reproducibility of ocular biometry and intraocular lens (IOL) power measurements obtained by ophthalmology residents using an AL-Scan device, a novel optical biometer. Methods. Two ophthalmology residents were instructed regarding the AL-Scan device. Both performed ocular biometry and IOL power measurements using AL-Scan, three times on each of 128 eyes,...
متن کاملPrecision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace
PURPOSE To evaluate the repeatability and reproducibility of corneal power measurements obtained by Topcon KR-1W and iTrace, and assess the agreement with measurements obtained by Allegro Topolyzer and IOLMaster. METHODS The right eyes of 100 normal subjects were prospectively scanned 3 times using all the 4 devices. Another observer performed additional 3 consecutive scans using the Topcon K...
متن کاملUrethral length measurement in women during sonographic urethrocystography - an analysis of repeatability and reproducibility.
UNLABELLED There has been a rise in the use of sonographic urethrocystography in patients with a full bladder. So far, no publications have been made on the analysis of repeatability and reproducibility of the measurements performed during this procedure. AIM An assessment of repeatability and reproducibility of urethral length measurements during sonographic urethrocystography in females wit...
متن کاملRepeatability and Reproducibility Standard Deviations in the Measurement of Trace Moisture Generated Using Permeation Tubes
Permeation-tube moisture generators are used in industry as calibrated sources of water vapor and carrier gas mixtures. Measurements were made using three permeation-tube moisture generators of the type used in the semiconductor industry. This paper describes repeatability and reproducibility standard deviations in measurement of moisture concentration from such generators. Repeatability refers...
متن کاملRepeatability and Reproducibility of Noninvasive Keratograph 5M Measurements in Patients with Dry Eye Disease
Purpose. To determine the intraexaminer repeatability and interexaminer reproducibility of tear meniscus height (TMH) and noninvasive Keratograph tear breakup time (NIKBUT) measurements obtained with the Keratograph 5M (K5M) in a sample of healthy and dry eye populations. Methods. Forty-two patients with dry eye disease (DED group) and 42 healthy subjects (healthy group) were recruited in this ...