Do Ceramic Femoral Heads Reduce Taper Fretting Corrosion in Hip Arthroplasty? A Retrieval Study
نویسندگان
چکیده
BACKGROUND Previous studies regarding modular head-neck taper corrosion were largely based on cobalt chrome (CoCr) alloy femoral heads. Less is known about head-neck taper corrosion with ceramic femoral heads. QUESTIONS/PURPOSES We asked (1) whether ceramic heads resulted in less taper corrosion than CoCr heads; (2) what device and patient factors influence taper fretting corrosion; and (3) whether the mechanism of taper fretting corrosion in ceramic heads differs from that in CoCr heads. METHODS One hundred femoral head-stem pairs were analyzed for evidence of fretting and corrosion using a visual scoring technique based on the severity and extent of fretting and corrosion damage observed at the taper. A matched cohort design was used in which 50 ceramic head-stem pairs were matched with 50 CoCr head-stem pairs based on implantation time, lateral offset, stem design, and flexural rigidity. RESULTS Fretting and corrosion scores were lower for the stems in the ceramic head cohort (p=0.03). Stem alloy (p=0.004) and lower stem flexural rigidity (Spearman's rho=-0.32, p=0.02) predicted stem fretting and corrosion damage in the ceramic head cohort but not in the metal head cohort. The mechanism of mechanically assisted crevice corrosion was similar in both cohorts although in the case of ceramic femoral heads, only one of the two surfaces (the male metal taper) engaged in the oxide abrasion and repassivation process. CONCLUSIONS The results suggest that by using a ceramic femoral head, CoCr fretting and corrosion from the modular head-neck taper may be mitigated but not eliminated. CLINICAL RELEVANCE The findings of this study support further study of the role of ceramic heads in potentially reducing femoral taper corrosion.
منابع مشابه
Fretting and Corrosion Damage at the Head-Neck Taper is Reduced with Ceramic Femoral Heads: A Retrieval Study
Introduction: Modular head-neck total hip arthroplasty (THA) offers several benefits for orthopaedic surgeons, including intraoperative flexibility and the ability to change the head at revision surgery [1]. However, modularity at the head-neck taper, especially the use of metallic heads and stems, can lead to fretting and corrosion at the taper junction, leading to metal ions and debris releas...
متن کاملRetrieval analysis of metal and ceramic femoral heads on a single CoCr stem design
OBJECTIVES The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromi...
متن کاملEffect of Taper Design on Trunnionosis in Metal on Polyethylene Total Hip Arthroplasty.
This study examines how taper design affects corrosion and fretting at the head trunnion surface. All hip prostheses retrieved between 1999 and 2013 with 28mm/+0 heads were selected, resulting in 44 cobalt-chrome-on-polyethylene implants, representing six taper designs. Mean implantation time: 8.9±3.7years. The femoral head tapers were scored for fretting and corrosion using the Goldberg scale ...
متن کاملSpontaneous modular femoral head dissociation complicating total hip arthroplasty.
Modular femoral heads have been used successfully for many years in total hip arthroplasty. Few complications have been reported for the modular Morse taper connection between the femoral head and trunnion of the stem in metal-on-polyethylene bearings. Although there has always been some concern over the potential for fretting, corrosion, and generation of particulate debris at the modular junc...
متن کاملFemoral taperosis: an accident waiting to happen?
A modular femoral head-neck junction has practical advantages in total hip replacement. Taper fretting and corrosion have so far been an infrequent cause of revision. The role of design and manufacturing variables continues to be debated. Over the past decade several changes in technology and clinical practice might result in an increase in clinically significant taper fretting and corrosion. T...