Joint design of large-tip-angle parallel RF pulses and blipped gradient trajectories.
نویسندگان
چکیده
PURPOSE To design multichannel large-tip-angle kT-points and spokes radiofrequency (RF) pulses and gradient waveforms for transmit field inhomogeneity compensation in high field magnetic resonance imaging. THEORY AND METHODS An algorithm to design RF subpulse weights and gradient blip areas is proposed to minimize a magnitude least-squares cost function that measures the difference between realized and desired state parameters in the spin domain, and penalizes integrated RF power. The minimization problem is solved iteratively with interleaved target phase updates, RF subpulse weights updates using the conjugate gradient method with optimal control-based derivatives, and gradient blip area updates using the conjugate gradient method. Two-channel parallel transmit simulations and experiments were conducted in phantoms and human subjects at 7 T to demonstrate the method and compare it to small-tip-angle-designed pulses and circularly polarized excitations. RESULTS The proposed algorithm designed more homogeneous and accurate 180° inversion and refocusing pulses than other methods. It also designed large-tip-angle pulses on multiple frequency bands with independent and joint phase relaxation. Pulses designed by the method improved specificity and contrast-to-noise ratio in a finger-tapping spin echo blood oxygen level dependent functional magnetic resonance imaging study, compared with circularly polarized mode refocusing. CONCLUSION A joint RF and gradient waveform design algorithm was proposed and validated to improve large-tip-angle inversion and refocusing at ultrahigh field.
منابع مشابه
Joint design of trajectory and RF pulses for parallel excitation.
We propose an alternating optimization framework for the joint design of excitation k-space trajectory and RF pulses for small-tip-angle parallel excitation. Using Bloch simulations, we show that compared with conventional designs with predetermined trajectories, joint designs can often excite target patterns with improved accuracy and reduced total integrated pulse power, particularly at high ...
متن کاملMultidimensional Arbitrary-Flip-Angle Parallel Transmit Pulse Design Using an Optimal Control Approach
INTRODUCTION The majority of the design methods for parallel transmit radiofrequency (RF) pulses so far are based on small-tip-angle (STA) approximation of the Bloch equation [13]. These methods can design only excitation pulses with a small flip angle (≤ 90°). We have proposed a noniterative method to design large-tip-angle (LTA) parallel transmit pulses (> 90°), based on an extension of the s...
متن کاملAdditive angle method for fast large-tip-angle RF pulse design in parallel excitation.
Current methods for parallel excitation RF pulse design are based on the small-tip-angle approximation, which provides a computationally efficient means of pulse calculation. In general, pulses designed with those methods are inaccurate when scaled to produce large-tip angles, and methods for large-tip-angle pulse design are more computationally demanding. This paper introduces a fast iterative...
متن کاملFast and Accurate Large-Tip-Angle RF Pulse Design for Parallel Excitation Using a Perturbation Analysis of the Bloch Equation
Introduction: Designing RF pulses in parallel excitation (PTX) using the linear small-tip-angle approximation [1-3] produces distorted excitation patterns at large tip angles due to the nonlinear nature of the Bloch equation. Grissom et al [4] proposed the ‘Additive Angle’ (AA) method for large-tip-angle RF pulse as a means to improve upon the small-tip-angle approximation (STA). This iterative...
متن کاملMagnetic Resonance in Medicine 59:779–787 (2008) Additive Angle Method for Fast Large-Tip-Angle RF Pulse Design in Parallel Excitation
Current methods for parallel excitation RF pulse design are based on the small-tip-angle approximation, which provides a computationally efficient means of pulse calculation. In general, pulses designed with those methods are inaccurate when scaled to produce large-tip angles, and methods for large-tipangle pulse design are more computationally demanding. This paper introduces a fast iterative ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Magnetic resonance in medicine
دوره 75 3 شماره
صفحات -
تاریخ انتشار 2016