Accuracy of 3D volumetric image registration based on CT, MR and PET/CT phantom experiments
نویسندگان
چکیده
Registration is critical for image-based treatment planning and image-guided treatment delivery. Although automatic registration is available, manual, visual-based image fusion using three orthogonal planar views (3P) is always employed clinically to verify and adjust an automatic registration result. However, the 3P fusion can be time consuming, observer dependent, as well as prone to errors, owing to the incomplete 3-dimensional (3D) volumetric image representations. It is also limited to single-pixel precision (the screen resolution). The 3D volumetric image registration (3DVIR) technique was developed to overcome these shortcomings. This technique introduces a 4th dimension in the registration criteria beyond the image volume, offering both visual and quantitative correlation of corresponding anatomic landmarks within the two registration images, facilitating a volumetric image alignment, and minimizing potential registration errors. The 3DVIR combines image classification in real-time to select and visualize a reliable anatomic landmark, rather than using all voxels for alignment. To determine the detection limit of the visual and quantitative 3DVIR criteria, slightly misaligned images were simulated and presented to eight clinical personnel for interpretation. Both of the criteria produce a detection limit of 0.1 mm and 0.1 degree. To determine the accuracy of the 3DVIR method, three imaging modalities (CT, MR and PET/CT) were used to acquire multiple phantom images with known spatial shifts. Lateral shifts were applied to these phantoms with displacement intervals of 5.0+/-0.1 mm. The accuracy of the 3DVIR technique was determined by comparing the image shifts determined through registration to the physical shifts made experimentally. The registration accuracy, together with precision, was found to be: 0.02+/-0.09 mm for CT/CT images, 0.03+/-0.07 mm for MR/MR images, and 0.03+/-0.35 mm for PET/CT images. This accuracy is consistent with the detection limit, suggesting an absence of detectable systematic error. This 3DVIR technique provides a superior alternative to the 3P fusion method for clinical applications.
منابع مشابه
A Novel Subsampling Method for 3D Multimodality Medical Image Registration Based on Mutual Information
Mutual information (MI) is a widely used similarity metric for multimodality image registration. However, it involves an extremely high computational time especially when it is applied to volume images. Moreover, its robustness is affected by existence of local maxima. The multi-resolution pyramid approaches have been proposed to speed up the registration process and increase the accuracy of th...
متن کاملA phantom study of the accuracy of CT, MR and PET image registrations with a block matching-based algorithm Étude sur fantôme de la précision d’un algorithme de recalage par blocs pour des images TDM, IRM et TEP
Purpose. – The aim of the present study was to quantitatively assess the performance of a block matching-based automatic registration algorithm integrated within the commercial treatment planning system designated ISOgrayTM from Dosisoft. The accuracy of the process was evaluated by a phantom study on computed tomography (CT), magnetic resonance (MR) and positron emission tomography (PET) image...
متن کاملComparison of state-of-the-art atlas-based bone segmentation approaches from brain MR images for MR-only radiation planning and PET/MR attenuation correction
Introduction: Magnetic Resonance (MR) imaging has emerged as a valuable tool in radiation treatment (RT) planning as well as Positron Emission Tomography (PET) imaging owing to its superior soft-tissue contrast. Due to the fact that there is no direct transformation from voxel intensity in MR images into electron density, itchr('39')s crucial to generate a pseudo-CT (Computed Tomography) image ...
متن کاملPhantom validation of coregistration of PET and CT for image-guided radiotherapy.
Radiotherapy treatment planning integrating positron emission tomography (PET) and computerized tomography (CT) is rapidly gaining acceptance in the clinical setting. Although hybrid systems are available, often the planning CT is acquired on a dedicated system separate from the PET scanner. A limiting factor to using PET data becomes the accuracy of the CT/PET registration. In this work, we us...
متن کاملNew Pseudo-CT Generation Approach from Magnetic Resonance Imaging using a Local Texture Descriptor
Background: One of the challenges of PET/MRI combined systems is to derive an attenuation map to correct the PET image. For that, the pseudo-CT image could be used to correct the attenuation. Until now, most existing scientific researches construct this pseudo-CT image using the registration techniques. However, these techniques suffer from the local minima of the non-rigid deformation energy f...