Dissolution study and method validation of alprazolam by high performance liquid chromatography method in pharmaceutical dosage form
ثبت نشده
چکیده
Dissolution study was carried for alprazolam solubility study, conditions selected for paddle was 100 rpm, using 500 ml potassium dihydrogen phosphate of pH 6.8 at temperature as 37°C. These conditions are useful for % release of alprazolam was achieved on symmetyshield RP8 (150 x 3.9 mm i.d, 5 μ.) column was most suitable for dissolution study with mixture of (60:40 % v/v) of buffer and acetonitrile as mobile phase. A and to tri-ethylamine was used to adjust the pH 3.0 of buffer. The diluent. The detector wavelength was set at 225 nm. performed as per ICH guidelines. The linearity was found between 50 RSD of precision was 0.66%.
منابع مشابه
Development and validation of spectrophotometric method for simultaneous estimation of paracetamol and lornoxicam in different dissolution media
BACKGROUND Paracetamol and lornoxicam in combined tablet dosage form are available in the market. This combination is used to treat inflammatory diseases of the joints, osteoarthritis and sciatica. Spectrophotometric and high performance liquid chromatography (HPLC) methods have been reported for their simultaneous estimation in tablet dosage form in specific solvent. This paper presents simple...
متن کاملDevelopment and validation of a selective online dissolution method for rosiglitazone maleate.
A new specific, accurate, precise, and reproducible selective online dissolution method for rosiglitazone maleate is developed and validated for the dissolution of rosiglitazone maleate in pharmaceutical formulations. The rationale of the method is based on the direct measurement of the absorbance of the analyte in the buffer medium at 242 nm using buffer as blank. Dissolution is achieved on a ...
متن کاملHigh performance liquid chromatographic method for determination of ezetimibe in pharmaceutical formulation tablets
Ezetimide belongs to a class of lipid lowering compounds that selectively inhibits intestinal absorption of cholesterol and related phytosterols. The purpose of this study is to establish a reliable and quick method for the assignment of ezetimibe in tablets form b...
متن کاملStability Indicating HPLC Method for the Determination of Fulvestrant in Pharmaceutical Formulation in Comparison with Linear Sweep Voltammetric Method
This paper describes two rapid, sensitive and specific methods for the determination of fulvestrant in pharmaceutical preparations by high performance liquid chromatography (HPLC) and linear sweep voltammetry (LSV). HPLC method was used to study the degradation behaviour. Fulvestrant was subjected to degradation under the conditions of hydrolysis (acid and alkali), oxidation (30% H2O2). The lin...
متن کاملStability Indicating HPLC Method for the Determination of Fulvestrant in Pharmaceutical Formulation in Comparison with Linear Sweep Voltammetric Method
This paper describes two rapid, sensitive and specific methods for the determination of fulvestrant in pharmaceutical preparations by high performance liquid chromatography (HPLC) and linear sweep voltammetry (LSV). HPLC method was used to study the degradation behaviour. Fulvestrant was subjected to degradation under the conditions of hydrolysis (acid and alkali), oxidation (30% H2O2). The lin...
متن کامل