Wear-pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total knee prostheses.
نویسندگان
چکیده
Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R(2) = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.
منابع مشابه
Wear Distribution Detection of Knee Joint Prostheses by Means of 3D Optical Scanners
The objective of this study was to examine total knee polyethylene inserts from in vitro simulation to evaluate and display-using a 3D optical scanner-wear patterns and wear rates of inserts exposed to wear by means of simulators. Various sets of tibial inserts have been reconstructed by using optical scanners. With this in mind, the wear behavior of fixed and mobile bearing polyethylene knee c...
متن کاملComparative gravimetric wear analysis in mobile versus fixed-bearing posterior stabilized total knee prostheses.
Polyethylene (PE) wear is the limiting factor for the longevity of a conventional total knee arthroplasty (TKA). Excessive wear leads to loosening and eventual implant failure. The aim of our in vitro study was to investigate wear of a PE tibial insert on a rotating platform as compared to the same insert fixed to the tibial baseplate and articulating with a similar femoral component. All tests...
متن کاملDifferences in long-term fixation between mobile-bearing and fixed-bearing knee prostheses at ten to 12 years' follow-up: a single-blinded randomised controlled radiostereometric trial.
This single-blinded randomised controlled trial investigated whether one design of mobile-bearing (MB) total knee replacement (TKR) has any advantage over a fixed-bearing (FB) design on long-term fixation as measured by radiostereometry. The amount of wear underneath the mobile bearing was also evaluated. A series of 42 knees was randomised to MB or FB tibial components with appropriate polyeth...
متن کاملA Low Conforming, Low Wear Solution in Fixed Bearing Total Knee Prostheses
Introduction: Ultra High Molecular Weight Polyethylene (PE) remains the primary bearing surface of choice in total knee replacements (TKR). Traditionally Archard’s Law is used to predict wear rates in hip and knee prostheses. This assumes that the volume of wear between two articulating surfaces is proportional to the applied load and sliding distance. However it has been shown that this is a s...
متن کاملIncrease in the Tibial Slope Reduces Wear after Medial Unicompartmental Fixed-Bearing Arthroplasty of the Knee
Introduction. Unicompartmental arthroplasty of the knee in patients with isolated medial osteoarthritis gives good results, but survival is inferior to that of total knee prosthesis. Knees may fail because positioning of the prosthesis has been suboptimal. The aim of this study was to investigate the influence of the tibial slope on the rate of wear of a medial fixed-bearing unicompartmental kn...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 92 4 شماره
صفحات -
تاریخ انتشار 2010