Flexible growing rods: a pilot study to determine if polymer rod constructs may provide stability to skeletally immature spines
نویسندگان
چکیده
BACKGROUND Surgical treatments for early onset scoliosis (EOS), including growing rod constructs, involve many complications. Some are due to biomechanical factors. A construct that is more flexible than current instrumentation systems may reduce complications. The purpose of this preliminary study was to determine spine range of motion (ROM) after implantation of simulated growing rod constructs with a range of clinically relevant structural properties. The hypothesis was that ROM of spines instrumented with polyetheretherketone (PEEK) rods would be greater than metal rods and lower than noninstrumented controls. Further, adjacent segment motion was expected to be lower with polymer rods compared to conventional systems. METHODS Biomechanical tests were conducted on 6 skeletally immature porcine thoracic spines (domestic swine, 35-40 kg). Spines were harvested after death from swine that had been utilized for other studies (IACUC approved) which had not involved the spine. Paired pedicle screws were used as anchors at proximal and distal levels. Specimens were tested under the following conditions: control, then dual rods of PEEK (6.25 mm), titanium (4 mm), and CoCr (5 mm) alloy. Lateral bending (LB) and flexion-extension (FE) moments of ±5 Nm were applied. Vertebral rotations were measured using video. Differences were determined by two-tailed t-tests and Bonferroni correction with four primary comparisons: PEEK vs control and PEEK vs CoCr, in LB and FE (α=0.05/4). RESULTS In LB, ROM of specimens with PEEK rods was lower than control at each instrumented level. ROM was greater for PEEK rods than both Ti and CoCr at every instrumented level. Mean ROM at proximal and distal noninstrumented levels was lower for PEEK than for Ti and CoCr. In FE, mean ROM at proximal and distal noninstrumented levels was lower for PEEK than for metal. Combining treated levels, in LB, ROM for PEEK rods was 35% of control (p<0.0001) and 270% of CoCr rods (p<0.01). In FE, ROM with PEEK was 27% of control (p<0.001) and 180% of CoCr (p<0.01). CONCLUSIONS PEEK rods decreased flexibility versus noninstumented controls, and increased flexibility versus metal rods. Smaller increases in ROM at proximal and distal adjacent motion segments occurred with PEEK compared to metal rods, which may help decrease junctional kyphosis. Flexible growing rods may eventually help improve treatment options for young patients with severe deformity.
منابع مشابه
Flexible growing rods: polymer rods provide stability to skeletally immature spines
Objective Surgical treatments for early onset scoliosis (EOS) typically require multiple operations and many complications. A more flexible growing rod construct might result in a more flexible spine with fewer complications. Polymer rods (polyetheretherketone, PEEK) are relatively flexible in bending, and so might allow for greater range of motion (ROM) during treatment. The purpose of this st...
متن کاملFlexible growing rods: a biomechanical pilot study of polymer rod constructs in the stability of skeletally immature spines
BACKGROUND Surgical treatments for early onset scoliosis (EOS) correct curvatures and improve respiratory function but involve many complications. A distractible, or 'growing rod,' implant construct that is more flexible than current metal rod systems may sufficiently correct curves in small children and reduce complications due to biomechanical factors. The purpose of this pilot study was to d...
متن کاملFlexible growing rods: polymer rods provide stability to -skeletally immature spines
Introduction: Surgical treatments for early onset scoliosis typically require multiple operations and many complications. A more fl exible growing rod construct might result in a more fl exible spine with fewer complications. Polymer rods (polyetheretherketone, PEEK) are relatively fl exible in bending, and therefore might allow for greater range of motion (ROM) during treatment. The purpose of...
متن کاملBiomechanical Analysis of Anterior Spinal Instrumentation Constructs for Lumbar Fusions
Corrective surgery for scoliosis with anterior instrumentation can be accomplished with different constructs: single rod, single rod with interbody cages and dual rods. These varying constructs are most often used in the lumbar spine to generate a fusion mass while retaining the natural lordosis of the lumbar region (Sweet, 1999). In a pediatric population or in smaller adults, two screws canno...
متن کاملReduction of intradiscal pressure by the use of polycarbonate-urethane rods as compared to titanium rods in posterior thoracolumbar spinal fixation
Loss of sagittal alignment and balance in adult spinal deformity can cause severe pain, disability and progressive neurological deficit. When conservative treatment has failed, spinal fusion using rigid instrumentation is currently the salvage treatment to stop further curve progression. However, fusion surgery is associated with high revision rates due to instrumentation failure and proximal j...
متن کامل