Fracture fixation with two locking screws versus three non-locking screws

نویسندگان

  • B. Grawe
  • T. Le
  • S. Williamson
  • A. Archdeacon
  • L. Zardiackas
چکیده

OBJECTIVES We aimed to further evaluate the biomechanical characteristics of two locking screws versus three standard bicortical screws in synthetic models of normal and osteoporotic bone. METHODS Synthetic tubular bone models representing normal bone density and osteoporotic bone density were used. Artificial fracture gaps of 1 cm were created in each specimen before fixation with one of two constructs: 1) two locking screws using a five-hole locking compression plate (LCP) plate; or 2) three non-locking screws with a seven-hole LCP plate across each side of the fracture gap. The stiffness, maximum displacement, mode of failure and number of cycles to failure were recorded under progressive cyclic torsional and eccentric axial loading. RESULTS Locking plates in normal bone survived 10% fewer cycles to failure during cyclic axial loading, but there was no significant difference in maximum displacement or failure load. Locking plates in osteoporotic bone showed less displacement (p = 0.02), but no significant difference in number of cycles to failure or failure load during cyclic axial loading (p = 0.46 and p = 0.25, respectively). Locking plates in normal bone had lower stiffness and torque during torsion testing (both p = 0.03), but there was no significant difference in rotation (angular displacement) (p = 0.84). Locking plates in osteoporotic bone showed lower torque and rotation (p = 0.008), but there was no significant difference in stiffness during torsion testing (p = 0.69). CONCLUSIONS The mechanical performance of locking plate constructs, using only two screws, is comparable to three non-locking screw constructs in osteoporotic bone. Normal bone loaded with either an axial or torsional moment showed slightly better performance with the non-locking construct.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

First Metatarsophalangeal Joint Arthrodesis: A Retrospective Comparison of Crossed-screws, Locking and Non-Locking Plate Fixation with Lag Screw

Background:Locking plate fixation is increasingly used for first metatarsophalangeal joint (MTP-I) arthrodesis. Still there is few comparable clinical data regarding this procedure. Methods:We retrospectively evaluated 60 patients who received an arthrodesis of the MTP-I between January 2008 and June 2010. With 20 patients each we performed a locking plate fixation with lag screw, arthrodesis w...

متن کامل

First Metatarsophalangeal Joint Arthrodesis: A Retrospective Comparison of Crossed-screws, Locking and Non-Locking Plate Fixation with Lag Screw

Background:Locking plate fixation is increasingly used for first metatarsophalangeal joint (MTP-I) arthrodesis. Still there is few comparable clinical data regarding this procedure. Methods:We retrospectively evaluated 60 patients who received an arthrodesis of the MTP-I between January 2008 and June 2010. With 20 patients each we performed a locking plate fixation with lag screw, arthrodesis w...

متن کامل

A Pitfall in Fixation of Distal Humeral Fractures with Pre-Contoured Locking Compression Plate

Anatomically precontoured locking plates are intended to facilitate the fixation of articular fractures and particularly those associated with osteoporosis. Fractures of the distal humerus are relatively uncommon injuries where operative intervention can be exceptionally challenging. The distal humeral trochlea provides a very narrow anatomical window through which to pass a fixed-angle locking...

متن کامل

Locked plate fixation of osteoporotic humeral shaft fractures: are two locking screws per segment enough?

OBJECTIVE The purpose of this study was to compare the biomechanical behavior of using two versus three locking screws per bone segment in a cadaveric humerus fracture gap model. METHODS Six matched pairs of elderly osteoporotic fresh-frozen human cadaveric humerii were used. An eight-hole locking compression plate was placed posteriorly on the humeral shaft and secured with either four or si...

متن کامل

Single column locking plate fixation is inadequate in two column acetabular fractures. A biomechanical analysis

BACKGROUND The objective of this study was to determine whether one can achieve stable fixation of a two column (transverse) acetabular fracture by only fixing a single column with a locking plate and unicortical locking screws. We hypothesized that a locking plate applied to the anterior column of a transverse acetabular fracture would create a construct that is more rigid than a non-locking p...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 1  شماره 

صفحات  -

تاریخ انتشار 2012