BACKGROUND Vascular closure devices (VCDs) are frequently used for hemostasis with endovascular procedures by employing sutures or plug devices (using collagen or hydrogel) or through the use of a metal clip made of nickel and titanium, such as the StarClose SE device. In comparison to manual compression (MC), VCDs are associated with earlier time to discharge and ambulation, improved patient comfort, and better cost-effectiveness. CASE REPORT A 77-year-old man with history of ischemic cardiomyopathy with non-ST segment elevation myocardial infarction (NSTEMI) underwent diagnostic cardiac catheterization with deployment of a StarClose SE vascular closure device for hemostasis. Upon repeat access 4 days later for coronary intervention, retrograde sheath angiography revealed a pseudo-aneurysm emanating from the center of the StarClose clip. CONCLUSIONS A review of the literature shows VCDs to be non-inferior to MC, with an overall high success rate. Major and minor complications rates are comparable to those with MC, and pseudo-aneurysm is an infrequent complication.
منابع مشابه
StarClose vascular closure device: safety and efficacy of deployment and reaccess in a neurointerventional radiology service.
BACKGROUND AND PURPOSE Obtaining safe and effective closure of the femoral access site following neurointerventional procedures can sometimes be challenging, especially in patients on anti-coagulation or anti-platelet therapy. The purpose of this study was to evaluate the safety and efficacy of a novel percutaneous closure device that employs a nitinol clip-mediated extravascular closure strate...
متن کاملEndoscopic repair of an injured internal carotid artery utilizing femoral endovascular closure devices.
OBJECTIVES/HYPOTHESIS Injury to the internal carotid artery is a feared complication of endoscopic endonasal surgery of the skull base. Such an event, although rare, is associated with high morbidity and mortality. Even if bleeding is controlled, permanent neurological defects frequently persist. Many techniques have been developed to manage internal carotid artery rupture with varying degrees ...
متن کاملSafety and Efficacy of the StarClose Vascular Closure System following 8-Fr Sheath Placement for Intra-Aortic Balloon Pump: A Single-Center Analysis of 42 Consecutive Patients
OBJECTIVE To assess the safety and efficacy of the StarClose device following intra-aortic balloon pump (IABP) counterpulsation using 8-Fr femoral sheaths. SUBJECTS AND METHODS From June 2008 to August 2012, 42 consecutive patients who received IABP implantation via common femoral artery (CFA) punctures with an 8-Fr sheath (which were then sealed with the StarClose Vascular Closure System at ...
متن کاملAcute limb ischemia caused by incorrect deployment of a clip-based arterial closure device
Failure of a vascular closure device most commonly results in a hemorrhage or pseudoaneurysm formation. In this paper a rare case of severe acute limb ischemia following incorrect deployment of a clip-based closure device (Starclose SE, Abbott Vascular) in a 31-year-old woman is presented. Symptoms of acute limb ischemia occurred at the start of the ambulation, 6 h after completion of the proce...
متن کاملStarclose SE® hemostasis after 6F direct antegrade superficial femoral artery access distal to the femoral head for peripheral endovascular procedures in obese patients.
PURPOSE Direct superficial femoral artery (SFA) antegrade puncture is a valid alternative to common femoral artery (CFA) access for peripheral vascular interventions. Data investigating vascular closure device (VCD) hemostasis of distant SFA 6F access are limited. We aimed to investigate the safety and effectiveness of the Starclose SE® VCD for hemostasis, following direct 6F antegrade SFA acce...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The American journal of case reports
دوره 17 شماره
صفحات -
تاریخ انتشار 2016