Decontamination of targeted pathogens from patient rooms using an automated ultraviolet-C-emitting device.
نویسندگان
چکیده
OBJECTIVE. To determine the effectiveness of an automated ultraviolet-C (UV-C) emitter against vancomycin-resistant enterococci (VRE), Clostridium difficile, and Acinetobacter spp. in patient rooms. DESIGN. Prospective cohort study. SETTING. Two tertiary care hospitals. PARTICIPANTS. Convenience sample of 39 patient rooms from which a patient infected or colonized with 1 of the 3 targeted pathogens had been discharged. INTERVENTION. Environmental sites were cultured before and after use of an automated UV-C-emitting device in targeted rooms but before standard terminal room disinfection by environmental services. RESULTS. In total, 142 samples were obtained from 27 rooms of patients who were colonized or infected with VRE, 77 samples were obtained from 10 rooms of patients with C. difficile infection, and 10 samples were obtained from 2 rooms of patients with infections due to Acinetobacter. Use of an automated UV-C-emitting device led to a significant decrease in the total number of colony-forming units (CFUs) of any type of organism (1.07 log10 reduction; P < .0001), CFUs of target pathogens (1.35 log10 reduction; P < .0001), VRE CFUs (1.68 log10 reduction; P < .0001), and C. difficile CFUs (1.16 log10 reduction; P < .0001). CFUs of Acinetobacter also decreased (1.71 log10 reduction), but the trend was not statistically significant (P = .25). CFUs were reduced at all 9 of the environmental sites tested. Reductions similarly occurred in direct and indirect line of sight. CONCLUSIONS. Our data confirm that automated UV-C-emitting devices can decrease the bioburden of important pathogens in real-world settings such as hospital rooms.
منابع مشابه
Evaluation of an automated ultraviolet radiation device for decontamination of Clostridium difficile and other healthcare-associated pathogens in hospital rooms
BACKGROUND Environmental surfaces play an important role in transmission of healthcare-associated pathogens. There is a need for new disinfection methods that are effective against Clostridium difficile spores, but also safe, rapid, and automated. METHODS The Tru-D Rapid Room Disinfection device is a mobile, fully-automated room decontamination technology that utilizes ultraviolet-C irradiati...
متن کاملEvaluation of a hand-held far-ultraviolet radiation device for decontamination of Clostridium difficile and other healthcare-associated pathogens
BACKGROUND Environmental surfaces play an important role in transmission of healthcare-associated pathogens. There is a need for new disinfection methods that are effective against Clostridium difficile spores, but also safe and rapid. The Sterilray™ Disinfection Wand device is a hand-held room decontamination technology that utilizes far-ultraviolet radiation (185-230 nm) to kill pathogens. ...
متن کاملFirst UK evaluation of an automated ultraviolet-C room decontamination device (Tru-D™).
Tru-D™ is an automated room disinfection device that uses ultraviolet-C radiation to kill micro-organisms. The device was deployed in six side-rooms and an operating theatre. In a cleaned, unoccupied operating theatre, Tru-D eradicated all organisms from the environment. Using artificially seeded Petri dishes with meticillin-resistant Staphylococcus aureus, multi-resistant acinetobacter and van...
متن کاملEvaluation of an automated room decontamination device using aerosolized peracetic acid.
Because manual cleaning is often suboptimal, there is increasing interest in use of automated devices for room decontamination. We demonstrated that an ultrasonic room fogging system that generates submicron droplets of peracetic acid and hydrogen peroxide eliminated Clostridium difficile spores and vegetative pathogens from exposed carriers in hospital rooms and adjacent bathrooms.
متن کاملTerminal decontamination of patient rooms using an automated mobile UV light unit.
OBJECTIVE To determine the ability of a mobile UV light unit to reduce bacterial contamination of environmental surfaces in patient rooms. METHODS An automated mobile UV light unit that emits UV-C light was placed in 25 patient rooms after patient discharge and operated using a 1- or 2-stage procedure. Aerobic colony counts were calculated for each of 5 standardized high-touch surfaces in the...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Infection control and hospital epidemiology
دوره 34 5 شماره
صفحات -
تاریخ انتشار 2013