Design, assembly, and optical bench testing of a high-numerical-aperture miniature injection-molded objective for fiber-optic confocal reflectance microscopy.

نویسندگان

  • Matthew D Chidley
  • Kristen D Carlson
  • Rebecca R Richards-Kortum
  • Michael R Descour
چکیده

The design, analysis, assembly methods, and optical-bench test results for a miniature injection-molded plastic objective lens used in a fiber-optic confocal reflectance microscope are presented. The five-lens plastic objective was tested as a stand-alone optical system before its integration into a confocal microscope for in vivo imaging of cells and tissue. Changing the spacing and rotation of the individual optical elements can compensate for fabrication inaccuracies and improve performance. The system performance of the miniature objective lens is measured by use of an industry-accepted slanted-edge modulation transfer function (MTF) metric. An estimated Strehl ratio of 0.61 and a MTF value of 0.66 at the fiber-optic bundle Nyquist frequency have been obtained. The optical bench testing system is configured to permit interactive optical alignment during testing to optimize performance. These results are part of an effort to demonstrate the manufacturability of low-cost, high-performance biomedical optics for high-resolution in vivo imaging. Disposable endoscopic microscope objectives could help in vivo confocal microscopy technology mature to permit wide-scale clinical screening and detection of early cancers and precancerous lesions.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

In vivo fiber-optic confocal reflectance microscope with an injection-molded plastic miniature objective lens.

For in vivo optical diagnostic technologies to be distributed to the developed and developing worlds, optical imaging systems must be constructed of inexpensive components. We present a fiber-optic confocal reflectance microscope with a cost-effective injection-molded plastic miniature objective lens for in vivo imaging of human tissues in near real time. The measured lateral resolution is less...

متن کامل

Design of a high-numerical-aperture miniature microscope objective for an endoscopic fiber confocal reflectance microscope.

An endoscopic confocal microscope requires a high-performance, miniaturized microscope objective. We present the design of a miniature water-immersion microscope objective that is approximately 10 times smaller in length than a typical commercial objective. The miniature objective is 7 mm in outer diameter and 21 mm in length (from object to image). It is used in a fiber confocal reflectance mi...

متن کامل

Design and demonstration of a miniature catheter for a confocal microendoscope.

The fluorescence confocal microendoscope provides high-resolution, in vivo imaging of cellular pathology during optical biopsy. The confocal microendoscope employs a flexible fiber-optic catheter coupled to a custom-built slit-scan confocal microscope. The catheter consists of a fiber-optic imaging bundle linked to a miniature objective and focus assembly. The 3-mm-diameter catheter may be used...

متن کامل

Slit-scanning confocal microendoscope for high-resolution in vivo imaging.

We discuss the design and construction of a novel imaging system in which a fiber-optic imaging bundle and miniature optical and mechanical components are used to allow confocal fluorescence microscopy in remote locations. The instrumentation has been developed specifically for cellular examination of tissue for optical biopsy. Miniaturization of various components makes the device usable in a ...

متن کامل

In-vivo multi-spectral confocal microscopy

A multi-spectral confocal microendoscope (MCME) for in-vivo imaging has been developed. The MCME employs a flexible fiber-optic catheter coupled to a slit-scan confocal microscope with an imaging spectrometer. The catheter consists of a fiber-optic imaging bundle linked to a miniature objective and focus assembly. The focus mechanism allows for imaging to a maximum tissue depth of 200 microns. ...

متن کامل