Static structural testing of trans-tibial composite sockets.

نویسندگان

  • T A Current
  • G F Kogler
  • D G Barth
چکیده

The purpose of this investigation was to quantify the structural strength of various trans-tibial composite sockets. To conduct the study, loading parameters and methods were developed that emulate the International Standards Organisation (ISO) standards for structural testing of lower limb prostheses since specific guidelines for the testing of the trans-tibial socket portion of a prosthesis have not yet been established. The experimental set-up simulated the instant of maximum loading during the late stance phase of gait. Ten trans-tibial sockets were evaluated. Five different reinforcement materials and two resin types were used to construct the sockets. A standard four hole distal attachment plate was used to connect the socket and pylon. Each sample was loaded to failure in a servo-hydraulic materials test machine at 100 N/s. None of the composites in the study met the ISO 10328 standards for level A100, loading condition II (4025 N), as required for other prosthetic componentry. All failures occurred at the site of the pyramid attachment plate. Ultimate strength and failure type were material dependent. Load point deflection was significantly different for the resin variable (p<0.05). Statistical differences according to reinforcement material were noted in composite weight and strength-to-weight ratio (p<0.05). The fibre volume fraction was also estimated and recorded. Reinforcement material type was the primary determinant of performance for the tested samples. Carbon reinforcements performed better than fibreglass reinforcements of similar weave type. The greatest ultimate strength and strength-to-weight ratio was observed with the unidirectional carbon reinforcement.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Computer-aided socket design for trans-femoral amputees.

Computer-aided socket design (CASD) is a technique for the design of prosthetic sockets using the advantageous features of computer graphics and calculation. This paper describes a computer-aided technique used to design sockets for patients with trans-tibial lower limb amputations.

متن کامل

Results of the pilot phase of a clinical evaluation of computer aided design of trans-tibial prosthesis sockets.

This study served as a pilot to develop the methodology for a larger number of clinical trials. Ten trans-tibial amputees compared sockets made by conventional methods with sockets made using computer aided socket design (CASD). Prosthetists, paired for experience, fitted each subject with one prosthesis using each method. A preferred socket was selected by the subject on the basis of comfort w...

متن کامل

Clinical evaluation of trans-tibial prosthesis sockets: a comparison between CAD CAM and conventionally produced sockets.

This study is an evaluation, from the patient's point of view, of CAD CAM prosthesis sockets compared with conventional sockets. Twenty-two trans-tibial amputees were divided into two groups. One group was provided with a CAD CAM (CAPOD) socket, the other with a conventionally made one. After one month the groups were evaluated with regard to subjective experience, the judgement of a prosthetis...

متن کامل

Analysis of trans tibial prosthetic socket materials using finite element method

The objective of this work was to analyze in a parametric study for optimum design solution of prosthetic socket material by finite element method. A realistic three-dimensional finite element model of the PTB socket was developed to find out the stress distribution pattern under physiologically relevant loading condition during normal walking. The CAD model of the rectified socket was collecte...

متن کامل

Sand-casting technique for trans-tibial prostheses.

A sand-casting technique for trans-tibial sockets was applied to 28 amputees, and the prosthetic fit observed through transparent check sockets. The results were better than historical reports of fittings with plaster of Paris casting by qualified prosthetists. The fit was consistently and evenly larger than the stump, but total contact could be achieved by applying three (two to five) stump so...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Prosthetics and orthotics international

دوره 23 2  شماره 

صفحات  -