Consistency of vendor-specified activity values for ¹⁹²Ir brachytherapy sources.
نویسندگان
چکیده
A long-term comparison was done between the manufacturer-stated ¹⁹²Ir activity and the measured ¹⁹²Ir activities determined with a well-type ionization chamber. Sources for a Nucletron Micro Selectron high-dose-rate (HDR) unit were used for this purpose. The radioactive sources reference activities were determined using a PTW well-type ionization chamber traceable to the National Institute of Standards and Technology Primary Calibration Laboratory. The measurements were taken in a period of 56 months with 17 different radioactive sources. The manufacturer stated activities were taken from the source calibration certificate provided by the manufacturer. These values were compared with the measured activities. The results have shown that both the percentage deviation of the monthly control measurements with the well-type chamber and the ratio between the measured activities to the manufacturer-stated value lie within ± 2.5%. These results were compared with similar published data and with uncertainty level (3% of the mean and 5% maximum deviation from mean) for brachytherapy sources calibration recommended by the AAPM. It was concluded that a threshold level of ±2.5% can be used as a suitable quality assurance indicator to spot problems in our department. The typical ±5% uncertainty as provided by the manufacturers may be tightened to ±3% to be more in line with published AAPM reports.
منابع مشابه
Estimation of distance error by fuzzy set theory required for strength determination of HDR 192Ir brachytherapy sources
Verification of the strength of high dose rate (HDR) (192)Ir brachytherapy sources on receipt from the vendor is an important component of institutional quality assurance program. Either reference air-kerma rate (RAKR) or air-kerma strength (AKS) is the recommended quantity to specify the strength of gamma-emitting brachytherapy sources. The use of Farmer-type cylindrical ionization chamber of ...
متن کاملCalculating of Dose Distribution in Tongue Brachytherapy by Different Radioisotopes using Monte Carlo Simulation and Comparing by Experimental Data
Introduction: Among different kinds of oral cavity cancers, the frequency of tongue cancer occurrence is more significant. Brachytherapy is the most common method to cure tongue cancers. Long sources are used in different techniques of tongue brachytherapy. The objective of this study is to asses the dose distribution around long sources, comparing different radioisotopes as brachytherapy sourc...
متن کاملUpdating the Planar Patterson-Parker Table for Ir-192 and Cs-137 Brachytherapy Sources Using the Most Recent TG-43U1 Recommended Dosimetric Parameters
Background: The Patterson-Parker table was created in 1934 to determine mg-hr required to deliver the prescribed dose to the treatment area. These tables were created using the dosimetric data for 226Ra that was determined in air and were utilized for dose calculations around implants with Ra-equivalent radionuclides such as 137Cs and 192Ir. Therefore, the differences of the tissue attenuation ...
متن کاملMaterial-specific Conversion Factors for Different Solid Phantoms Used in the Dosimetry of Different Brachytherapy Sources
Introduction Based on Task Group No. 43 (TG-43U1) recommendations, water phantom is proposed as a reference phantom for the dosimetry of brachytherapy sources. The experimental determination of TG-43 parameters is usually performed in water-equivalent solid phantoms. The purpose of this study was to determine the conversion factors for equalizing solid phantoms to water. Materials and Methods T...
متن کاملSize Effects of Gold and Iron Nanoparticles on Radiation Dose Enhancement in Brachytherapy and Teletherapy: A Monte Carlo Study
Introduction In this study, we aimed to calculate dose enhancement factor (DEF) for gold (Au) and iron (Fe) nanoparticles (NPs) in brachytherapy and teletherapy, using Monte Carlo (MC) method. Materials and Methods In this study, a new algorithm was introduced to calculate dose enhancement by AuNPs and FeNPs for Iridium-192 (Ir-192) brachytherapy and Cobalt-60 (Co-60) teletherapy sources, using...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical dosimetry : official journal of the American Association of Medical Dosimetrists
دوره 37 1 شماره
صفحات -
تاریخ انتشار 2012