All-polyethylene versus metal-backed and stemmed tibial components in cemented total knee arthroplasty. A prospective, randomised RSA study.
نویسندگان
چکیده
We studied the quality of fixation of the tibial component using radiostereometric analysis (RSA) in 40 patients who had undergone a cemented Freeman-Samuelson total knee arthroplasty. They were prospectively randomised to either a stemmed metal-backed (MB) or non-stemmed all-polyethylene (AP) tibial component. The articulating geometry of the implants was identical, as was the operative technique and the postoperative regime. The study showed no complications of fixation using AP tibial components, and the migration was the same as that of their metal-backed counterparts. There was no bony collapse or increased subsidence of any part of the tibial component or increased incidence of radiolucent lines in the knees with AP components. Most AP implants were stable between one and two years after surgery, a finding known to be of positive prognostic significance when predicting future aseptic loosening.
منابع مشابه
Total knee arthroplasty Aspects on improved fixation in the younger patient Av
Mobile bearings do not improve fixation in cemented total knee arthroplasty. Trabecular metal tibial component in patients younger than 60 years. A prospective RSA study. Cemented versus uncemented fixation of the femoral component of the NexGen CR total knee replacement in patients younger than 60 years. A randomized RSA study. Submitted 2008.
متن کاملThe all-polyethylene tibial component in primary total knee arthroplasty.
Outcomes of total knee arthroplasties performed with modern all-polyethylene tibial components have been found to be comparable with or better than those of arthroplasties done with metal-backed modular components in numerous mid-to-long-term follow-up studies, radiostereometric analyses, and the few prospective randomized trials available. Advantages of an all-polyethylene tibial component ove...
متن کاملAll-Polyethylene Versus Metal-Backed Tibial Components-An Analysis of 27,733 Cruciate-Retaining Total Knee Replacements from the Swedish Knee Arthroplasty Register.
BACKGROUND Currently, the use of metal-backed tibial components is more common than the use of all-polyethylene components in total knee arthroplasty. However, the available literature indicates that all-polyethylene tibial components are not inferior to the metal-backed design. We hypothesized that there would be no difference in the ten-year survival rate between all-polyethylene and metal-ba...
متن کاملSurvivorship of cemented knee replacements.
The survivorship method of analysis has been used to compare the failure rate and overall success of 1,430 cemented primary total knee arthroplasties performed at The Hospital for Special Surgery over a 15-year period. There were 224 total condylar prostheses with a polyethylene tibia, 289 of the posterior stabilised type with an all polyethylene tibia, and 917 posterior stabilised with a metal...
متن کاملMetal-backed versus all-polyethylene unicompartmental knee arthroplasty
OBJECTIVES Up to 40% of unicompartmental knee arthroplasty (UKA) revisions are performed for unexplained pain which may be caused by elevated proximal tibial bone strain. This study investigates the effect of tibial component metal backing and polyethylene thickness on bone strain in a cemented fixed-bearing medial UKA using a finite element model (FEM) validated experimentally by digital image...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 83 6 شماره
صفحات -
تاریخ انتشار 2001