Accuracy of noninvasive multiwave pulse oximetry compared with carboxyhemoglobin from blood gas analysis in unselected emergency department patients.

نویسندگان

  • Dominik Roth
  • Harald Herkner
  • Wolfgang Schreiber
  • Nina Hubmann
  • Gunnar Gamper
  • Anton N Laggner
  • Christof Havel
چکیده

STUDY OBJECTIVE Accurate and timely diagnosis of carbon monoxide (CO) poisoning is difficult because of nonspecific symptoms. Multiwave pulse oximetry might facilitate the screening for occult poisoning by noninvasive measurement of carboxyhemoglobin (COHb), but its reliability is still unknown. We assess bias and precision of COHb oximetry compared with the criterion standard blood gas analysis. METHODS This was a prospective diagnostic accuracy study according to STARD (Standards for the Reporting of Diagnostic accuracy studies) criteria, performed at a tertiary care hospital emergency department. We included all patients for whom both invasive and noninvasive measurement within 60 minutes was available, regardless of their complaints, during a 1-year period. RESULTS One thousand five hundred seventy-eight subjects were studied, of whom 17 (1.1%) received a diagnosis of CO poisoning. In accordance with this limited patient cohort, we found a bias of 2.99% COHb (1.50% for smokers, 4.33% for nonsmokers) and a precision of 3.27% COHb (2.90% for smokers, 2.98% for nonsmokers), limits of agreement from -3.55% to 9.53% COHb (-4.30% to 7.30% for smokers, -1.63% to 10.29% for nonsmokers). Upper limit of normal cutoff of 6.6% COHb had the highest sensitivity in screening for CO poisoning. Smoking status and COHb level had the most influence on the deviation between measurements. CONCLUSION Multiwave pulse oximetry was found to measure COHb with an acceptable bias and precision. These results suggest it can be used to screen large numbers of patients for occult CO poisoning.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Emergency department management of suspected carbon monoxide poisoning: role of pulse CO-oximetry.

BACKGROUND The RAD-57 pulse CO-oximeter is a lightweight device allowing noninvasive measurement of blood carboxyhemoglobin (S(pCO)). We assessed the diagnostic value of pulse CO-oximetry, comparing S(pCO) values from the RAD-57 to standard laboratory blood carboxyhemoglobin (COHb) measurement in emergency department patients with suspected carbon monoxide (CO) poisoning. METHODS This was a p...

متن کامل

Can we Replace Arterial Blood Gas Analysis by Pulse Oximetry in Neonates with Respiratory Distress Syndrome, who are Treated According to INSURE Protocol?

Neonates with respiratory distress syndrome (RDS), who are treated according to INSURE protocol; require arterial blood gas (ABG) analysis to decide on appropriate management. We conducted this study to investigate the validity of pulse oximetry instead of frequent ABG analysis in the evaluation of these patients. From a total of 193 blood samples obtained from 30 neonates <1500 grams with RDS,...

متن کامل

Reliability of new pulse CO-oximeter in victims of carbon monoxide poisoning.

STUDY OBJECTIVE The purpose of this study was to evaluate the reliability of noninvasive real-time measurement of carboxyhemoglobin (COHb) using a pulse CO-oximeter in victims of carbon monoxide poisoning (COP). METHODS During the 7-month study period, pulse CO-oximetry was measured on patients admitted to the emergency department (ED) for suspected COP. Each patient included in the study und...

متن کامل

A new tool for the early diagnosis of carbon monoxide intoxication.

Invasive measurement of carboxyhemoglobin (COHb) by blood gas analysis (BGA) is accepted as the standard diagnostic procedure in diagnosis of inhalation injury and carbon monoxide (CO) intoxications. The main disadvantage of BGA with COHb testing is the unavailability in pre-hospital rescue conditions. The non-invasive SpCO analysis using pulse CO oximetry (Rad57, Masimo Corp., USA) represents ...

متن کامل

The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study

BACKGROUND Pulse oximetry is routinely used to continuously and noninvasively monitor arterial oxygen saturation (SaO2) in critically ill patients. Although pulse oximeter oxygen saturation (SpO2) has been studied in several patient populations, including the critically ill, its accuracy has never been studied in emergency department (ED) patients with severe sepsis and septic shock. Sepsis res...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Annals of emergency medicine

دوره 58 1  شماره 

صفحات  -

تاریخ انتشار 2011