Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55.

نویسندگان

  • Y C Cheung
  • K N Yu
  • R T Ho
  • C P Yu
چکیده

The accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (model-B) radiosurgery was verified using both the GafChromic films MD-55 (improved) and the EGS4 Monte Carlo technique. The Monte Carlo technique was employed to calculate the dose distribution along the x-, y- and z-axes when a single shot with opening of all 201 sources was delivered at the centre of a simulated water phantom with a diameter of 160 mm. Collimator helmets with different size of the Gamma Knife unit were verified. Good consistency (typical discrepancy less than 2%) was obtained between the results of Monte Carlo and GammaPlan. Small discrepancies, however, were obtained by GafChromic films. Discrepancies, as great as 10% when using the 4 mm collimator helmet, at the low percentage isodose curve along the z-axis of the measurement results were probably due to the small energy dependency of the GafChromic films. Significant discrepancies were not observed along x- and y-axes because such small discrepancies were easily over-washed by other gamma beams coming from the x- and y-directions. Similar results showing the discrepancies between the GafChromic films and GammaPlan were obtained when using the 8, 14 and 18 mm collimator helmets. However, the discrepancies along the z-axis became smaller as scattering effect increased when using larger collimator sizes. We suggest that the Monte Carlo technique should also be applied in stereotactic dose planning system verification as it is an ideal and reliable computational technique.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte Carlo calculations and GafChromic film measurements for plugged collimator helmets of Leksell Gamma Knife unit.

The Monte Carlo technique and GafChromic films were employed to verify the accuracy of the dose planning system (Leksell GammaPlan) used in Gamma Knife (type B) radiosurgery when plugged collimator helmets were used. The EGS4 Monte Carlo code was used to calculate the dose distribution along the x, y, and z axes when a single shot was delivered at the center point (unit center point: x = 100, y...

متن کامل

Dose distributions at extreme irradiation depths of gamma knife radiosurgery: EGS4 Monte Carlo calculations.

The accuracy of the dose planning system (Leksell GammaPlan), used in Gamma Knife (type B) radiosurgery at extreme irradiation depths, was verified using the Monte Carlo technique. EGS4 Monte Carlo calculations were employed to calculate the dose distribution along the x, y and z axes for an irradiation relatively shallow in a spherical bony cavity water phantom. Two different sizes of the coll...

متن کامل

Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife

Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with eac...

متن کامل

Study of scattered photons from the collimator system of Leksell Gamma Knife using the EGS4 Monte Carlo code.

In the algorithm of Leksell GAMMAPLAN (the treatment planning software of Leksell Gamma Knife), scattered photons from the collimator system are presumed to have negligible effects on the Gamma Knife dosimetry. In this study, we used the EGS4 Monte Carlo (MC) technique to study the scattered photons coming out of the single beam channel of Leksell Gamma Knife. The PRESTA (Parameter Reduced Elec...

متن کامل

Monte Carlo calculation of single-beam dose profiles used in a gamma knife treatment planning system.

The accuracy of single-beam dose profiles used in the algorithm of the Gamma Knife treatment planning system (Leksell GammaPlan) is verified. EGS4 Monte Carlo calculation was employed to calculate the dose distributions of single-beams in a spherical water phantom with diameter 160 mm. The beams were directed to the center of the phantom. Collimators of 4, 8, 14, and 18 mm sizes were studied. T...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine

دوره 53 3  شماره 

صفحات  -

تاریخ انتشار 2000