Safety of localizing epilepsy monitoring intracranial electroencephalograph electrodes using MRI: Radiofrequency-induced heating
نویسندگان
چکیده
PURPOSE To investigate heating during postimplantation localization of intracranial electroencephalograph (EEG) electrodes by MRI. MATERIALS AND METHODS A phantom patient with a realistic arrangement of electrodes was used to simulate tissue heating during MRI. Measurements were performed using 1.5 Tesla (T) and 3T MRI scanners, using head- and body-transmit RF-coils. Two electrode-lead configurations were assessed: a "standard" condition with external electrode-leads physically separated and a "fault" condition with all lead terminations electrically shorted. RESULTS Using a head-transmit-receive coil and a 2.4 W/kg head-average specific absorption rate (SAR) sequence, at 1.5T the maximum temperature change remained within safe limits (<1 degrees C). Under "standard" conditions, we observed greater heating (<or=2.0 degrees C) at 3T on one system and similar heating (<1 degrees C) on a second, compared with the 1.5T system. In all cases these temperature maxima occurred at the grid electrode. In the "fault" condition, larger temperature increases were observed at both field strengths, particularly for the depth electrodes. Conversely, with a body-transmit coil at 3T significant heating (+6.4 degrees C) was observed (same sequence, 1.2/0.5 W/kg head/body-average) at the grid electrode under "standard" conditions, substantially exceeding safe limits. These temperature increases neglect perfusion, a major source of heat dissipation in vivo. CONCLUSION MRI for intracranial electrode localization can be performed safely at both 1.5T and 3T provided a head-transmit coil is used, electrode leads are separated, and scanner-reported SARs are limited as determined in advance for specific scanner models, RF coils and implant arrangements. Neglecting these restrictions may result in tissue injury.
منابع مشابه
Feasibility of simultaneous intracranial EEG - fMRI : a safety study
In epilepsy patients who have electrodes implanted in their brains as part of their pre-surgical assessment, simultaneous intracranial EEG and fMRI (icEEG-fMRI) may provide important localising information and improve understanding of the underlying neuropathology. However, patient safety during icEEG-fMRI has not been addressed. Here the potential health hazards associated with icEEG-fMRI were...
متن کاملMRI-related heating near deep brain stimulation electrodes: more data are needed.
Magnetic resonance imaging (MRI) of patients with implanted deep brain stimulation (DBS) devices poses a challenge for healthcare providers. As a consequence of safety concerns about magnetic field interactions with the device, induced electrical currents and thermal damage due to radiofrequency heating, a number of stringent guidelines have been proposed by the device manufacturer. Very few de...
متن کاملPTFOS: Flexible and Absorbable Intracranial Electrodes for Magnetic Resonance Imaging
Intracranial electrocortical recording and stimulation can provide unique knowledge about functional brain anatomy in patients undergoing brain surgery. This approach is commonly used in the treatment of medically refractory epilepsy. However, it can be very difficult to integrate the results of cortical recordings with other brain mapping modalities, particularly functional magnetic resonance ...
متن کاملComprehensive RF safety concept for parallel transmission MR.
PURPOSE The goal of this study is to increase patient safety in parallel transmission (pTx) MRI systems. A major concern in these systems is radiofrequency-induced tissue heating, which can be avoided by specific absorption rate (SAR) prediction and SAR monitoring before and during the scan. METHODS In this novel comprehensive safety concept, the SAR is predicted prior to the scan based on pr...
متن کاملLocalization of neurosurgically implanted electrodes via photograph-MRI-radiograph coregistration.
Intracranial electroencephalography (iEEG) is clinically indicated for medically refractory epilepsy and is a promising approach for developing neural prosthetics. These recordings also provide valuable data for cognitive neuroscience research. Accurate localization of iEEG electrodes is essential for evaluating specific brain regions underlying the electrodes that indicate normal or pathologic...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 28 شماره
صفحات -
تاریخ انتشار 2008