Perforation of the aortic sinus after closure of atrial septal defects with the Atriasept occluder.
نویسندگان
چکیده
Percutaneous atrial septal defect closure is routinely performed nowadays because of the ease of implantation as well as the low complication rate. The Atriasept ASD occluder is a low profile, double disc device; over the years several modifications have been made. We report two cases of aortic sinus perforation by the Atriasept ASD occluder (model 2007). Two asymptomatic patients, in whom the device was implanted, were noticed to have metal projecting into the aorta. Real-time fluoroscopy showed fractures of the outer metal ring with abnormal movement of one of the struts of the device. One patient is being conservatively managed and in the other the device was surgically removed due to the presence of a second ASD, which needed closure. Transesophageal echocardiography and fluoroscopy may be necessary to identify this potentially life-threatening complication of this device.
منابع مشابه
Penetration of left and right atrial wall and aortic root by an Amplatzer atrial septal occluder in a nine year old boy with Marfan syndrome: Case report
BACKGROUND To describe complications associated with Amplatzer septal occluders in a patient with Marfan syndrome CASE PRESENTATION A nine-year-old boy with Marfan syndrome and a 22 mm atrial septal defect (ASD) was treated successfully by interventional closure of his ASD by placing a 24 mm Amplatzer septal occluder. Follow up examinations showed a good result but an increasing enlargement o...
متن کاملUrgent Surgical Intervention for Embolized Cardiac Occluder Devices: A Case Series
Introduction: In this study, we sought to illustrate our experience in urgent surgical management for embolized cardiac septal occlude devices resulting from trans-catheter closure of atrial septal defect and ventricular septal defect. Mathrials and Methods: We retrospectively reviewed four patients aged 2–10 years who underwent urgent surgery due to cardiac septal occluder embolization between...
متن کاملCardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder.
OBJECTIVES Amplatzer septal occluder (ASO)-associated cardiac perforation (CP) at our institution prompted this retrospective review. BACKGROUND Cardiac perforation is a rare complication after transcatheter atrial septal defect (ASD) closure. METHODS To identify CP after transcatheter ASD closure with ASO, cardiac events (CE) describing definite CP, hemopericardium, pericardial effusion, c...
متن کاملEvaluation of Clinical Course in Children and Adolescents with Atrial Septal Defects
Background Atrial septal defects (ASDs) are the most common third congenital heart defects. This study aimed to evaluate the clinical course of ASDs and the relationship between its complications, location closure and size. Materials and Methods: This cross-sectional study was conducted...
متن کاملSurgical removal of fractured ATRIASEPT occluder 5 months after successful percutaneous closure of atrial septal defect.
A 10-year-old girl with symptoms of atypical infection and diagnosed disintegration of the occluder (Fig. 1) was referred for surgery. Intra-operatively the device was incorporated with atrial septum, but its surface was not completely endothelialized. All parts were removed (Fig. 2). The defect was closed with pericardial patch. www.elsevier.com/locate/ejcts European Journal of Cardio-thoracic...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
دوره 74 2 شماره
صفحات -
تاریخ انتشار 2009