Ventilation difficulty due to herniation of inflating tube of the cuff inside the endotracheal tube.
نویسندگان
چکیده
The complications due to airway equipments remain to be one of the most challenging situations for the anesthesiologists. Any part of an endotracheal tube may cause a risk when it leads to an obstruction in the airway. Here, we present an increased airway pressure due to herniation of inflating tube of the cuff inside an armored (wire reinforced) endotracheal tube during neurosurgery in prone position. A 34 years old, female patient was scheduled for 6th thoracic vertebraectomy and bilateral transpedinculer fixation after vertebrae fracture and dislocation. The patient was intubated uneventfully with an armoured tube (I.D. 7.5 mm, O.D. 17 mm, Mallinckrodt, St. Louis, USA). The tube level was checked by auscultation and taped at 19 cm. The initial airway pressure was between 14 16 cm H2O with a tidal volume of 450 ml and 11 breaths/min. The anesthesia was maintained by isoflurane and nitrous oxide/O2 (50%/50%). The endotracheal tube checked again for an unintentional malpositioning of the tube after prone positioning and confirmed by auscultation and airway pressure which remained between 18 and 20 cm H2O. After two hours, air leakage was suspected from the tube and the cuff was inflated with 1 ml of air and the leakage was ended. Airway pressure started to rise during rod placement after transpedinculer screws insertion at the 3rd hour, and the neurosurgeons were informed about the problem assuming that the rise in airway pressure might be related to the surgical procedure as the manipulations due to the reduction for obtaining thoracic lordosis could be the cause of increased airway pressure. We tried to aspirate through the endotracheal tube with a 14 Ch aspiration catheter. However, we couldn’t insert the catheter more than 5 cm inside the tube. The manual ventilation was started until suturation since the airway pressure and PaCO2 peaked at 60 cm H2O and 70 mmHg, respectively. After a rapid closure of the surgical site, the patient positioned supine, however airway pressure still remained high and the suctioning was impossible. When we inspected inside the tube, we noticed a transparent projection obstructing 1⁄2 of the inner diameter of the endotracheal tube (Fig. 1). Immediately, we removed the obstructed tube and replaced it with a standard 7.5 mm tube. The airway pressure returned into normal and increase in PaCO2 resolved quickly. The patient was extubated uneventfully.
منابع مشابه
Comparison of tracheal tube cuff inflation pressure among anesthesia team with different gradual level
Background & Objectives: Tracheal intubation is frequently performed in operating room. Over inflation of the endotracheal tube cuff is a risk factor for tracheal ischemia and subsequent complications. On the other hand inadequate cuff pressure increases the risk of aspiration of gastric contents. In this study, the endotracheal tube cuff pressure filled by the anesthesia team has been compared...
متن کاملRepeated Tracheostomy Tube Cuff Rupture Due to Tracheobronchopathia Osteochondroplastica: A Case Report
Introduction: Tracheobronchopathia osteochondroplastica (TPO) is a rare benign disorder of the lower part of the trachea and the upper part of the main bronchi. Case Report: A case of tracheobronchopathia osteochondroplastica (TPO) diagnosed at the time of intubation in an intensive care unit due to difficulty when advancing the endotracheal tube beyond the vocal cords, is reported. A problem...
متن کاملبررسی تاثیر تغییر وضعیت بدن بر فشار کاف لوله تراشه در بیماران تحت تهویه مکانیکی
Background: Endotracheal tube cuff pressure must be maintained in safe range. Many factors could be affecting on endotracheal tube cuff pressure in patients on mechanical ventilation. Patients undergoing mechanical ventilation (MV) in critical care settings require changing position for different reasons. The aim of this study was to determine the effect of changes in body position and head of ...
متن کاملStudy of endotracheal tube cuff pressure and time of measurement among intensive care units patients
Introduction: The pressure in the endotracheal tube cuff must be kept within an optimal range, to make sure tracheal capillaries circulation still stable during positive ventilation. This is essential to avoid oral and gastric stuffing aspiration Objective: The objective of this study was to assess the endotracheal tube cuff pressure and time intervals measurements among intensive care units...
متن کاملEffect of body position and head of bed angle on intra-abdominal and endotracheal tube cuff pressure in mechanically ventilated patients
Introduction: Changing position is one of the typical nursing care in intensive care units. This study aimed to investigate the effect of body positioning on intra-abdominal pressure and endotracheal tube cuff pressure and their relationship with each other in patients undergoing mechanical ventilation Materials and Methods: This was a before-after clinical trial study performed on 70 patients ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Middle East journal of anaesthesiology
دوره 21 3 شماره
صفحات -
تاریخ انتشار 2011