Development of a Computerized 4-D MRI Phantom for Liver Motion Study

نویسندگان

  • Chunhao Wang
  • Fang-Fang Yin
  • W. P. Segars
  • Zheng Chang
  • Lei Ren
چکیده

PURPOSE To develop a 4-dimensional computerized magnetic resonance imaging phantom with image textures extracted from real patient scans for liver motion studies. METHODS The proposed phantom was developed based on the current version of 4-dimensional extended cardiac-torso computerized phantom and a clinical magnetic resonance scan. Initially, the extended cardiac-torso phantom was voxelized in abdominal-chest region at the end of exhalation phase. Structures/tissues were classified into 4 categories: (1) Seven key textured organs, including liver, gallbladder, spleen, stomach, heart, kidneys, and pancreas, were mapped from a clinical T1-weighted liver magnetic resonance scan using deformable registration. (2) Large textured soft tissue volumes were simulated via an iterative pattern generation method using the same magnetic resonance scan. (3) Lung and intestine structures were generated by assigning uniform intensity with proper noise modeling. (4) Bony structures were generated by assigning the magnetic resonance values. A spherical hypointensity tumor was inserted into the liver. Other respiratory phases of the 4-dimensional phantom were generated using the backward deformation vector fields exported by the extended cardiac-torso program, except that bony structures were generated separately for each phase. A weighted image filtering process was utilized to improve the overall tissue smoothness at each phase. RESULTS Three 4-dimensional series with different motion amplitudes were generated. The developed motion phantom produced good illustrations of abdominal-chest region with anatomical structures in key organs and texture patterns in large soft tissue volumes. In a standard series, the tumor volume was measured as 13.90 ± 0.11 cm3 in a respiratory cycle and the tumor's maximum center-of-mass shift was 2.95 cm/1.84 cm on superior-inferior/anterior-posterior directions. The organ motion during the respiratory cycle was well rendered. The developed motion phantom has the flexibility of motion pattern variation, organ geometry variation, and tumor modeling variation. CONCLUSIONS A 4-D computerized phantom was developed and could be used to produce image series with synthetic magnetic resonance textures for magnetic resonance imaging research of liver motion.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Development of a Liver Phantom Based on Computed Tomography Images for Dosimetric Purpose

Introduction: The present study was conducted with the aim of designing a liver phantom for dosimetry. To benchmark the results obtained by the developed liver phantom, another method was applied for the dosimetry of a real liver tissue using imaging. Materials and Methods: For the purpose of the study, a real liver tissue was converted into a phantom based on thegram-molecular weight of the co...

متن کامل

Design and Fabrication of a Four-Dimensional Respiratory Phantom for Studying Tumor Movement in Radiotherapy with Magnetic Resonance Imaging

Introduction: In radiation therapy, determining the location of the tumor accurately during irradiation is one of the most important requirements. However, lung tumors are not fixed in a single location and move during irradiation due to respiratory motion. Due to limitations in assessing such movements, using a lung phantom can be useful and operational for their fast, easy an...

متن کامل

Medical imaging modalities: Prevention of unnecessary orders and non-optimized radiation exposure

Magnetic resonance imaging (MRI) uses a constant magnetic field and radio waves, a non-invasive method for examining tissues, organs and the skeletal system. Advantages and disadvantages of MRI are not fully understood; however, it has been commonly used in the diagnosis of skeleton and bone diseases (such as disk and spinal disorders, joint disorders (arthritis), bone tumours and tissues...

متن کامل

Comparison of functional MRI image realignment tools using a computer-generated phantom.

This study discusses the development of a computer-generated phantom to compare the effects of image realignment programs on functional MRI (fMRI) pixel activation. The phantom is a whole-head MRI volume with added random noise, activation, and motion. It allows simulation of realistic head motions with controlled areas of activation. Without motion, the phantom shows the effects of realignment...

متن کامل

Development of a 4-D digital mouse phantom for molecular imaging research.

PURPOSE We develop a realistic and flexible 4-D digital mouse phantom and investigate its usefulness in molecular imaging research. METHODS Organ shapes were modeled with non-uniform rational B-spline (NURBS) surfaces based on high-resolution 3-D magnetic resonance microscopy (MRM) data. Cardiac and respiratory motions were modeled based on gated magnetic resonance imaging (MRI) data obtained...