Double-spin-echo diffusion weighting with a modified eddy current adjustment.
نویسنده
چکیده
Magnetic field inhomogeneities like eddy current-related gradient fields cause geometric distortions in echo-planar imaging (EPI). This in particular affects diffusion-weighted imaging where these distortions vary with the direction of the diffusion weighting and hamper the accurate determination of diffusion parameters. The double-spin-echo preparation often used aims to reduce the cumulative eddy current effect by adjusting the diffusion-weighting gradient pulse durations to the time constant of the dominant eddy current contribution. However, eddy currents with a variety of time constants may be present and cause residual distortions. Here, a modification is proposed where the two bipolar gradient pairs of the preparation are adjusted independently to different time constants. At the expense of a slightly prolonged echo time, residual geometric distortions and correspondingly increased values of the diffusion anisotropy can be reduced as is demonstrated in phantoms and the human brain. Thus, it may help to improve the reliability of diffusion-weighted EPI.
منابع مشابه
Echo Planar Diffusion-weighted Mri of the Abdomen and Pelvis: Comparison of Free- Breathing Monopolar and Bipolar Spin-echo Sequences in Image Quality and Geometric Distortions
Introduction: Diffusion-weighted magnetic resonance imaging (DWI) exploits the random, thermally induced displacement of water molecules in biological tissue. Most DWI applications in the abdomen and pelvis employ single-shot spin-echo echo-planar imaging (ssEPI) due to its fast acquisition times and resultant robust performance against motion artifacts. Nevertheless EPI-based DWI suffers from ...
متن کاملAn assessment of eddy current sensitivity and correction in single-shot diffusion-weighted imaging.
Artefacts caused by eddy currents are a major problem in diffusion weighted imaging. This is particularly acute in experiments in which a number of images with differing degrees of diffusion weighting and/or differently oriented diffusion-weighting gradients need to be combined. The echo-planar imaging sequence is particularly sensitive to the effects of residual eddy currents, especially due t...
متن کاملPerformance of single spin-echo and doubly refocused diffusion-weighted sequences in the presence of eddy current fields with multiple components.
Echo-planar diffusion-weighted images can display significant geometric distortions due to eddy current fields. Several preparation schemes have been proposed, which can null eddy currents with a single time constant. The aim of this work was to compare the performance of three such pulse sequences in the presence of multiple components and investigate whether affine registration is capable of ...
متن کاملImproved convection compensating pulsed field gradient spin-echo and stimulated-echo methods.
The need for convection compensating methods in NMR has been manifested through an increasing number of publications related to the subject over the past few years (J. Magn. Reson. 125, 372 (1997); 132, 13 (1998); 131, 126 (1998); 118, 50 (1996); 133, 379 (1998)). When performing measurements at elevated temperature, small convection currents may give rise to erroneous values of the diffusion c...
متن کاملGradient preemphasis calibration in diffusion-weighted echo-planar imaging.
This article describes a method which enables fast and objective pulse-sequence-specific preemphasis calibration, using standard pulse sequences and system hardware. The method is based on a k-space measurement technique, and has been applied to single-shot, diffusion-weighted, spin-echo, echo-planar imaging (DW-SE-EPI), which is particularly sensitive to eddy-current-induced image distortions....