AIM To evaluate the cyclic fatigue resistance of Reciproc(®) and WaveOne(®) instruments in simulated root canals. METHODOLOGY Two groups of 15 NiTi endodontic instruments of identical tip size of 0.25 mm were tested, group A; Reciproc(®) R25 and group B: WaveOne(®) primary. Cyclic fatigue testing was performed in a stainless steel artificial canal manufactured by reproducing the instrument's size and taper. A simulated root canal with a 60° angle of curvature and 5-mm radius of curvature was constructed for both the instruments tested. The centre of the curvature was 5 mm from the tip of the instrument and the curved segment of the canal was approximately 5 mm in length. The Reciproc(®) instruments were activated using the preset programme specific for the Reciproc(®) instruments, whilst the WaveOne(®) instruments were activated using the preset programme specific for the WaveOne(®) instruments. All instruments were rotated until fracture occurred and the time to fracture (TtF) and the length of the fractured tip were recorded and registered. Means and standard deviations of TtF and fragment length were calculated for each system and data were subjected to Student's t-test (P < 0.05). RESULTS A statistically significant difference (P < 0.05) was noted between Reciproc(®) and WaveOne(®) instruments. Reciproc(®) R25 instruments were associated with a significant increase in the mean time to fracture when compared with primary WaveOne(®) instruments (130.8 ± 18.4 vs. 97.8 ± 15.9 s). There was no significant difference (P > 0.05) in the mean length of the fractured fragments between the instruments. CONCLUSIONS Reciproc(®) instruments were associated with a significantly higher cyclic fatigue resistance than WaveOne(®) instruments.
منابع مشابه
Influence of continuous or reciprocating motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments.
INTRODUCTION The aim of this study was to evaluate the resistance to flexural fatigue of Reciproc R25 (VDW, Munich, Germany), WaveOne Primary (VDW, Munich, Germany), Mtwo (Sweden & Martina, Padova, Italy), and Twisted File (TF; SybronEndo, Orange, CA) instruments used in continuous rotation or in 2 different reciprocating motions. METHODS A total of 180 nickel-titanium files from 4 brands mar...
متن کاملDifferences in cyclic fatigue resistance at apical and coronal levels of Reciproc and WaveOne new files.
INTRODUCTION Recent improvements in alloys, kinematics, and concepts have been combined to increase the cyclic fatigue resistance of nickel-titanium (NiTi) instruments. The aim of this study was to compare the cyclic fatigue resistance of new M-Wire reciprocating WaveOne (Dentsply Maillefer, Ballaigues, Switzerland) and Reciproc (VDW GmbH, Munich, Germany) files at 2 levels. METHODS Sixty Rec...
متن کاملComparison of cyclic fatigue and torsional resistance in reciprocating single-file systems and continuous rotary instrumentation systems.
As compared with continuous rotary systems, reciprocating motion is believed to increase the fatigue resistance of NiTi instruments. We compared the cyclic fatigue and torsional resistance of reciprocating single-file systems and continuous rotary instrumentation systems in simulated root canals. Eighty instruments from the ProTaper Universal, WaveOne, MTwo, and Reciproc systems (n = 20) were s...
متن کاملCyclic fatigue and torsional resistance of two new nickel-titanium instruments used in reciprocation motion: Reciproc versus WaveOne.
INTRODUCTION The use of reciprocating movement was claimed to increase the resistance of nickel-titanium (NiTi) file to fatigue in comparison with continuous rotation. Recently 2 new brands of NiTi files have been marketed for use in a RM mode. The purpose of this study was to compare the cyclic fatigue resistance and torsional resistance of these 2 files, Reciproc and WaveOne. METHODS Cyclic...
متن کاملA comparison of two reciprocating instruments using bending stress and cyclic fatigue tests.
The aim of this study was to comparatively evaluate the bending resistance at 45º, the static and dynamic cyclic fatigue life, and the fracture type of the WaveOne (Dentsply Maillefer, Ballaigues, Switzerland) 25-08 and Reciproc (VDW, Munich, Germany) 25-08 instruments. A total of 60 nickel-titanium (NiTi) instruments (30 Reciproc and 30 WaveOne) from three different lots, each of which was 25 ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- International endodontic journal
دوره 45 7 شماره
صفحات -
تاریخ انتشار 2012