Fracture of the Anterior Locking Flange of a Total Knee Arthroplasty Polyethylene Liner Presenting with Pain following Knee Replacement
نویسندگان
چکیده
INTRODUCTION Fracture of the modern polyethylene insert of a total knee arthroplasty is rare. We describe the first case of a fractured anterior locking flange of the commonly used Depuy Press-fit Condylar (PFC) Sigma prosthesis. CASE REPORT The 80 year old Caucasian gentleman presented 8 years following previously uncomplicated and successful primary total knee replacement with pain, swelling and symptoms of instability of the knee. He was able to sublux his knee posteriorly using his hamstrings. Dissociation of the liner was evident on radiographs. He underwent revision of the polyethylene liner. It was evident during the revision that the anterior locking flange of the polyethylene liner had fractured allowing it to dissociate from the tibial tray. At 12 months following this revision he continues to do well and has similar range of movement and function to prior to the episode. This cause of the failure is not clear. CONCLUSION Surgeons should be aware of this rare complication when assessing a painful or unstable total knee replacement.
منابع مشابه
Fracture of an uncemented tantalum patellar component
A 62-year-old man presented with the acute, atraumatic onset of pain 3 years after uncemented right total knee arthroplasty. He complained of new mechanical locking with the knee held in extension on examination and unable to flex the knee. On the plain radiographs, the patellar component peg was fractured and the plate was dislocated. The knee was immobilized, and revision to a cemented 3-peg ...
متن کاملStress Fracture of the Lateral Femoral Condyle afterTotal Knee Arthroplasty
Background: This is to assess one of the rare complications after total knee replacement and to assess risk factors of failure.Methods: 11 patients with varus knee and an average age of 67 years underwent TKA between 2005 and 2013. All patients returned with a sudden sharp knee pain, disability to walk and significant decrease in ROM about 4 to 8 weeks after surgery. Radiographic examination re...
متن کاملPolyethylene Tibial Post Fracture in a Posterior Stabilized (Genesis II) Total Knee Arthroplasty
Posterior stabilizing knee prostheses have been increasing in popularity recently due to some studies showing excellent long-term survivorship, more consistent kinematics, technical ease, and higher range of motion.1,2 The posterior stabilized knee relies on a cam and post mechanism to replicate native femoral rollback. One of the major potential complications of this system is post fracture. T...
متن کاملThe all-polyethylene tibial component in primary total knee arthroplasty.
Outcomes of total knee arthroplasties performed with modern all-polyethylene tibial components have been found to be comparable with or better than those of arthroplasties done with metal-backed modular components in numerous mid-to-long-term follow-up studies, radiostereometric analyses, and the few prospective randomized trials available. Advantages of an all-polyethylene tibial component ove...
متن کاملPolyethylene in knee arthroplasty: A review.
Polyethylene (PE) has been used extensively in knee arthroplasty since the mid 20th century. Progress in material manufacturing and processing has led to newer polyethylenes over last few decades with different material properties. It has been established that PE wear in knee arthroplasty causes particle induced osteolysis which is the main reason for late failure and requires revision surgery....
متن کامل