Optimization of urinary dipstick pH: Are multiple dipstick pH readings reliably comparable to commercial 24-hour urinary pH?
نویسندگان
چکیده
PURPOSE Accurate measurement of pH is necessary to guide medical management of nephrolithiasis. Urinary dipsticks offer a convenient method to measure pH, but prior studies have only assessed the accuracy of a single, spot dipstick. Given the known diurnal variation in pH, a single dipstick pH is unlikely to reflect the average daily urinary pH. Our goal was to determine whether multiple dipstick pH readings would be reliably comparable to pH from a 24-hour urine analysis. MATERIALS AND METHODS Kidney stone patients undergoing a 24-hour urine collection were enrolled and took images of dipsticks from their first 3 voids concurrently with the 24-hour collection. Images were sent to and read by a study investigator. The individual and mean pH from the dipsticks were compared to the 24-hour urine pH and considered to be accurate if the dipstick readings were within 0.5 of the 24-hour urine pH. The Bland-Altman test of agreement was used to further compare dipstick pH relative to 24-hour urine pH. RESULTS Fifty-nine percent of patients had mean urinary pH values within 0.5 pH units of their 24-hour urine pH. Bland-Altman analysis showed a mean difference between dipstick pH and 24-hour urine pH of -0.22, with an upper limit of agreement of 1.02 (95% confidence interval [CI], 0.45-1.59) and a lower limit of agreement of -1.47 (95% CI, -2.04 to -0.90). CONCLUSIONS We concluded that urinary dipstick based pH measurement lacks the precision required to guide medical management of nephrolithiasis and physicians should use 24-hour urine analysis to base their metabolic therapy.
منابع مشابه
A new device for simple and accurate urinary pH testing by the Stone-former patient
PURPOSE Urinary pH is an important factor linked to renal stone disease and a useful marker in the treatment of urolithiasis. Although the gold standard for measuring urinary pH utilizes a glass electrode and a pH meter, at present dipstick testing is largely used to estimate urinary pH. However, the accuracy and precision of this method may be insufficient for making clinical decisions in pati...
متن کاملRE: Dipstick Spot urine pH does not accurately represent 24 hour urine PH measured by an electrode
OBJECTIVES To determine whether spot urine pH measured by dipstick is an accurate representation of 24 hours urine pH measured by an electrode. MATERIALS AND METHODS We retrospectively reviewed urine pH results of patients who presented to the urology stone clinic. For each patient we recorded the most recente pH result measured by dipstick from a spot urine sample that preceded the result of...
متن کاملTHE VALUE OF DIPSTICK ANALYSIS OF URINARY PROTEIN IN PREGNANCY INDUCED HYPERTENSION
In order to determine the value of dipstick analysis of urinary protein in pregnancy induced hypertension, a prospective study analyzing pregnant patients with a diagnosis of hypertensive disorder was conducted to compare the result of urinary protein dipstick analysis with 24hr. urine protein collection in obstetrical clinics affiliated to Shiraz University of Medical Sciences. All patien...
متن کاملبررسی میزان شیوع باکتریوری بدون علامت در دانش آموزان دختر دبستانی یکی از شهرکهای اطراف تهران
Background & objectives: Urinary Tract Infection (UTI) which unfortunately has no clear symptoms in many cases is the second prevalent infection among children after respiratory infection. This study carried out in order to determine the frequency of asymptomatic UTI among elementary school girls in one of Tehran’s suburbs (Mahalat town). Methods: In a cross sectional study, all elementary ...
متن کاملDevelopment of Contrast-Induced Acute Kidney Injury after Elective Contrast Media Exposure in Patients with Type 2 Diabetes Mellitus: Effect of Albuminuria
BACKGROUND The influence of albuminuria and urinary pH on the development of contrast-induced acute kidney disease (CI-AKI) in patients with type 2 diabetes mellitus (T2DM) after elective coronary angiography (CAG) or percutaneous coronary intervention (PCI) is unknown. METHODS CI-AKI was defined as an increase in serum creatinine >26.4 µmol/L or ≥50% of baseline value within 48 hours after c...